Oxygen sensors (table 1), 4 english – Hach-Lange ORBISPHERE 31xxx Basic User Manual User Manual

Page 4

Advertising
background image

Specification

Membrane

2956A

Membrane

2952A

Membrane

2995A

Membrane

29015A

Response time

1

2 seconds

5 seconds

6 seconds

50 seconds

Recommended minimum liquid flow
rate

2

[mL/min]

50 to 220

40 to 200

20 to 70

20 to 40

Recommended minimum linear flow
rate

2

[cm/sec]

200

150

50

30

Recommended gaseous flow rate
[L/min]

0.005 to 3

1

Response time at 25°C for a 90% signal change

2

Liquid flow through an ORBISPHERE 32001 flow chamber, with protection cap and no grille

Oxygen sensors (table 1)

Specification

Membrane

2956A

Membrane

2958A

Membrane

29552A

Membrane

2952A

Recommended applications

Corrosion

control, De-

aerated water

Beverage, Lab.

applications

In line wort,

Air/O

2

injection,

Sewage

treatment

Corrosion

control, In line

beverage, De-

aerated water

Material

PFA

Tefzel

®

PTFE

Tefzel

®

Thickness [μm]

25

12.5

50

25

Calibration gas

Air

Air

Air

Air / pure O

2

Dissolved measurement range

0 ppb to 20 ppm 0 ppb to 40 ppm 0 ppb to 80 ppm 0 ppb to 80 ppm

Gaseous measurement range

0 Pa to 50 kPa

0 Pa to 100 kPa 0 Pa to 200 kPa 0 Pa to 200 kPa

Accuracy

The greater of

±1% of reading

or ± 0.1 ppb

(1)

,

or ± 1 ppb

(2)

, or

± 0.25 Pa

The greater of

±1% of reading
or ± 1 ppb, or ±

2 Pa

The greater of

±1% of reading
or ± 2 ppb, or ±

5 Pa

The greater of

±1% of reading
or ± 2 ppb, or ±

5 Pa

(1)

Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments

(2)

Accuracy is ± 1 ppb for 366x and 3650 instruments

Integrated radiation dose limit

2 x 10

4

10

8

N/A

10

8

Expected current in air @ 1 bar 25°C
[μA]

26.4

9.4

6.3

5.4

Expected current in pure O

2

[μA]

132

47

31.4

27

O

2

consumption in O

2

saturated water

at 25°C [μg/hour]

40

14

9.4

8

Temperature compensation range

-5 to 60°C

Temperature measuring range

-5 to 100° C

Response time

1

7.2 seconds

9.5 seconds

90 seconds

38 seconds

Recommended minimum liquid flow
rate

2

[mL/min]

180

120

50

50

4 English