Sensor membrane specifications, In-line instrument applications (table 1), In-line instrument applications (table 2) – Hach-Lange ORBISPHERE 31xxx Thermal Basic User Manual User Manual

Page 4: 4 english

Advertising
background image

Sensor membrane specifications

In-line instrument applications (Table 1)

Sensor type

31 250 H

2

and 31260 H

2

(purge gas N

2

)

For other purge gases, please contact your Hach Lange

representative for specification.

Specification

Membrane 29561A

Membrane 2952A

Membrane 2935A

Thickness [μm]

25

25

25

Material

PFA

ETFE

ECTFE (Halar

®

)

Recommended applications

Waste gas offgas,

reactor coolant

Reactor coolant

High H

2

level

Radiation limits

10

5

rad

10

8

rad

10

8

rad

Measurement range at 25°C

0-2 ppm, or

0-25 cc/kg, or

0-1.5 bar

0-10 ppm, or

0-120 cc/kg, or 0-6 bar

0-20 ppm, or

0-220 cc/kg, or

0-12 bar

Accuracy (sample temperature
20-50°C within ± 5°C of calibration
temperature)

The greater of ±1% of

reading or ± 2 ppb, or

± 0.03cc/kg, or ±

1.5 mbar

The greater of ±1% of

reading or ± 8 ppb, or

± 0.1cc/kg, or ±

6 mbar

The greater of ±1% of

reading or ± 25 ppb, or

± 0.4 cc/kg, or ±

20 mbar

Accuracy (sample temperature 0-50°C
independent of calibration
temperature)

The greater of ±3% of

reading or ± 15 ppb or

± 0.18 cc/kg or ±

6 mbar

The greater of ±3% of

reading or ± 60 ppb or

± 0.6 cc/kg, or ±

20 mbar

The greater of ±3% of

reading or ± 150 ppb

or ± 2.5 cc/kg, or ±

50 mbar

Measurement cycle (seconds)

17

Sample flow rate

1

(through flow

chamber)

220 mL/min

200 mL/min

100 mL/min

Linear flow rate

2

(past sensor socket)

N/A

N/A

N/A

1

Minimal flow through an ORBISPHERE 32001 flow chamber

2

Minimal flow past an ORBISPHERE 29501 sensor socket

In-line instrument applications (Table 2)

Sensor type

31 450 CO

2

and

31460 CO

2

(purge

gas N

2

)

31 550 N

2

and

31560 N

2

(purge

gas CO

2

)

31 280 H

2

/D

2

and He (clean air

purge)

Specification

Membrane 29561A

Membrane 29561A

Membrane

2952A

Membrane

29561A

Thickness [μm]

25

25

25

25

Material

PFA

PFA

ETFE

PFA

Recommended applications

In-line beverage

In-line beverage

Reactor coolant

Waste gas PWR

Radiation limits

10

5

10

5

10

8

10

5

Measurement range at 25°C

0-7 V/V, or

0-15 g/kg, or

0-10 bar

0-350 ppm, or

0-300 mL/L, or

0-20 bar

0-100% H

2

/D

2

or

0-10 bar

0-100% H

2

/D

2

or

0-4 bar

4 English

Advertising