Sensor weight and dimensions, General information – Hach-Lange ORBISPHERE 31xxx Thermal Basic User Manual User Manual

Page 5

Advertising
background image

Sensor type

31 450 CO

2

and

31460 CO

2

(purge

gas N

2

)

31 550 N

2

and

31560 N

2

(purge

gas CO

2

)

31 280 H

2

/D

2

and He (clean air

purge)

Specification

Membrane 29561A

Membrane 29561A

Membrane

2952A

Membrane

29561A

Accuracy (sample
temperature 20-50°C within
± 5°C of calibration
temperature)

The greater of ±1%

of reading or ±

0.006V/V, or ±

0.012g/kg, or ±

8 mbar

The greater of ±2%

of reading or ±

0.3 ppm, or ±

0.25mL/L, or ±

15 mbar

The greater of

±2% of reading

or ± 1% H

2

/D

2

The greater of

±2% of reading

or ± 0.2% H

2

/D

2

Accuracy (sample
temperature 0-50°C
independent of calibration
temperature)

The greater of ±2%

of reading or ±

0.02 V/V or ±

0.048g/kg or ±

14 mbar

The greater of ±4%

of reading or ±

1 ppm or ± 0.8 mL/L

or ± 34 mbar

N/A

N/A

Measurement cycle
(seconds)

22

1

60

60

Sample flow rate

2

(through

flow chamber)

100 mL/min

300 mL/min

200 mL/min

500 mL/min

Linear flow rate

3

(past

sensor socket)

50 cm/sec

150 cm/sec

N/A

N/A

1

Measurement cycle for 3624 ProBrix application is 15 seconds

2

Minimal flow through an ORBISPHERE 32001 flow chamber

3

Minimal flow past an ORBISPHERE 29501 sensor socket

Sensor weight and dimensions

Weight with external temperature adapter 1330g (±20g). Weight without external temperature
adapter 1040g (±20g).

Figure 1 Sensor dimensions

Note: Shown with external temperature adapter.

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.