Stablcal calibration procedure – Hach-Lange 2100 N Laboratory Turbidimeter Basic User Manual User Manual

Page 11

Advertising
background image

StablCal calibration procedure

1. Remove the filter
assembly. Refer to

Change the filter
assembly

on page 15.

2. Clean the lens of the
USEPA filter assembly.
Refer to

Clean the filter

assembly

on page 15.

3. Hold the tab of the
USEPA filter assembly
so that the arrows point
toward the front of the
instrument. Push the
filter assembly fully in
the housing.

4. Push CAL.
The S0 light turns on.
The NTU value of the
dilution water used in
the last calibration is
shown on the display.

5. Get the < 0.1 NTU
vial. Clean the vial with
a soft, lint-free cloth to
remove water spots and
fingerprints. Do not
invert the vial.

6. Apply a small bead
of silicone oil from the
top to the bottom of the
vial.

7. Use the oiling cloth
to apply the oil equally
to the surface of the
vial. Remove the
excess oil. Make sure
that the vial is almost
dry.

8. Put the vial in the
sample cell holder with
the triangle on the vial
aligned with the
reference mark on the
sample cell holder.
Close the cover.

9. Push ENTER.
The instrument display
counts down, then
measures the standard.
The next expected
standard (e.g., 20.00) is
shown. The S1 light
turns on.

10. Remove the vial
from the sample cell
holder.

11. Do steps 5–10 for
the other StablCal vials
(from lowest to highest
NTU standard).
The S0 light turns on
after the last vial is
measured.

12. Push CAL.
The instrument saves
the new calibration data
and goes back to
Measurement mode.

English 11

Advertising