Hach-Lange 2100 N Laboratory Turbidimeter Basic User Manual User Manual

Page 3

Advertising
background image

Table of contents

Specifications

on page 3

Standard operation

on page 9

General information

on page 4

Maintenance

on page 14

User interface

on page 7

Troubleshooting

on page 17

Startup

on page 9

Specifications

Specifications are subject to change without notice.

Specification

Details

Measurement method

Nephelometric

Regulatory

Meets EPA Method 180.1
ASTM D7315 - Standard Test Method for
Determination of Turbidity Above 1 Turbidity Unit (TU)
in Static Mode
ASTM D6655 - Standard Test Method for
Determination of Turbidity Below 5 NTU in Static Mode

Light source

Tungsten filament lamp

Measurement modes

NTU, NEPH (Nephelo) and EBC

Specification

Details

Range

NTU (Ratio on, manual range): 0–0.999, 0–9.99, 0–
99.9, 0–4000
NTU (Ratio on, auto range): 0–4000 auto decimal
NTU (Ratio off): 0–40.0
Nephelo (Ratio on, manual range): 0–9.99, 0–99.9, 0–
26,800
Nephelo (Ratio on, auto range): 0–26,800 auto
decimal
Nephelo (Ratio off): 0–268
EBC (Ratio on, manual range): 0–0.999, 0–9.99, 0–
99.9, 0–980
EBC (Ratio on, auto range): 0–980 auto decimal
EBC (Ratio off): 0–9.8

Accuracy

1,

2

Ratio on: ±2% of reading plus 0.01 NTU from 0–
1000 NTU, ±5% of reading from 1000–4000 NTU
based on formazin primary standard
Ratio off: ±2% of reading plus 0.01 NTU from 0–
40 NTU (under reference conditions

3

)

Resolution

Turbidity: 0.001 NTU/EBC, Nephelo (on lowest range)

Repeatability

±1% of reading or 0.01 NTU, whichever is greater
(under reference conditions)

Response time

Signal averaging off: 6.8 seconds
Signal averaging on: 14 seconds (when
10 measurements are used to calculate the average)

Stabilization time

Ratio on: 30 minutes after start-up
Ratio off: 60 minutes after start-up

Reading modes

Manual or auto range, signal averaging on or off, Ratio
on or off

Power requirement

115–230 VAC, 50/60 Hz (automatic power selection)
28 W maximum

English 3