Stablcal® rapidcal™ calibration, Stablcal, Rapidcal – Hach-Lange 2100Q and 2100Qis User Manual User Manual
Page 17: Calibration
Advertising
StablCal
®
RapidCal
™
calibration
1. Push the
CALIBRATION key to
enter the Calibration
mode. Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
2. Push the UP and
DOWN key to access
Cal Options and then
select Cal.Curve.
3. Select StablCal
®
RapidCal
™
from the list
and push OK.
4. Insert the 20 NTU
StablCal Standard and
close the lid.
Note: The standard to
be inserted is bordered.
5. Push Read. The
display shows
Stabilizing and then
shows the result.
6. Push Done to
review the calibration
details.
7. Push Store to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode,
refer to
Verification
calibration
.
English 15