:PULSe:PERiod

SOURce:PULSe:PERiod < period> sets the period of the pulse(s) to be
generated by the Pacer.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

pweiod

numeric

500E-9 to 8.388607 or

MIN| MAX

second

Comments

•

The resolution of period is 500E-9 seconds.

•

The Pacer waveform is a square wave with the output high for the first
half of the period, and low for the final half.

•

Related Commands: SOUR:PULS:COUN, ABORT, INIT,TRIG

•

*RST Condition: SOUR:PULS:PER 1E-6

Example

Setting the Pacer pulse period.

TRIG:SOUR HOLD

trigger source is TRIG
command

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

SOUR:PULS:PER .1 S

pulse period set to .1 second

INIT

go to Wait For Trigger state

TRIG

trigger the Pacer to output
pulses

:PULSe:PERiod?

SOURce:PULSe:PERiod? [ MIN | MAX] returns :

•

The current period if no parameter is sent.

•

The maximum allowable period if MAX is sent.

•

The minimum allowable period if MIN is sent.

Example

Querying the Pacer pulse period.

SOUR:PULS:PER?

ask for pulse period

enter statement

statement to enter value of
period

[SOURce] :PULSe:PERiod