Agilent Technologies E1300B User Manual

Page 180

Advertising
background image

[:IMMediate]

TRIGger:IMMediate will cause a trigger cycle to occur immediately, provided
that the trigger system has been initiated (INITiate).

Comments

•

Related Commands: ABORt, INITiate

•

*RST Condition: This command is an event and has no *RST condition.

Example

Triggering the Pacer.

TRIG:SOUR HOLD

trigger source is TRIG
command

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

SOUR:PULS:PER .1 S

pulse period set to .1 second

TRIG:DELAY .75 S

start Pacer .75 sec after trigger

INIT

go to Wait For Trigger state

TRIG

trigger Pacer to output pulses.

:SLOPe

TRIGger:SLOPe < slope> is for SCPI compatibility. The mainframe’s "Event
In" signal only triggers on a negative going edge.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

slope

discrete

NEGative

none

Comments

•

Trying to set

…

SLOPe to other than NEG will generate an error.

•

Related Commands: ABORt, INITiate,

:SLOPe?

TRIGger:SLOPe? returns the current trigger slope setting. Since the
mainframe’s "Event In" signal only triggers on a negative going edge,
TRIG:SLOP? will always return "NEG".

:SOURce

TRIGger:SOURce < trig_source> configures the trigger system to respond to
the specified source.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

trig_source

character

BUS| EXT| HOLD| IMM

none

TRIGger[:IMMediate]

7-52 System Instrument Command Reference

Advertising
This manual is related to the following products: