FIGURE 6: 3A SANITARY TURBINE FLOWMETER
DISASSEMBLY FOR 3A SANITARY METERS
REASSEMBLY FOR 3A SANITARY METERS
1. Remove O-ring.
2. Push entire assembly out through the O-ring side of the meter.
3. Review all parts for damage
• Check bearings thoroughly
• Check O-rings for cracks and nicks
NOTE: If any part appears damaged, DO NOT REASSEMBLE. Call the factory for instructions.
NOTE: Internals must go back together the exact way that they come out.
1. Assemble the internal assembly as illustrated.
2. Push the assembly into the housing on the O-ring side.
3. Butt clip hanger up against the step inside the housing
4. Reinstall the O-ring.
5. Make sure the rotor spins freely and the clip hangers do not turn at all.
ITEM DESCRIPTION
QTY
1
HOUSING
1
2
ROTOR
1
3
CONE
2
4
HANGER
2
5
O-RING
1
6
SHAFT (Carbide or Teflon)
1
Pg.11