SET-UP OPERATING MODE
SP1200 will alternate these 2 displays until “RESET” is pressed.
Correct product must be displayed in the upper left hand corner.
To change product, press “
←” key.
Press “CLEAR/FEED” key.
Enter product number desired:
With correct product displayed, check in upper right hand corner of the display for a temperature reading or the
words “No T/C”. If “No T/C” is displayed, press the “TC TOGGLE” key to switch on the temperature
compensation.
**All prover runs are to be made using temperature compensation.
**If “+426H” is displayed, the temperature probe or cable is open.
**If “-270.9L” is displayed, the temperature probe or cable is shorted.
PROVING MODE
LIN
Gm 0
No T/C
GO PROVE
00:00P
00-Mar-00
Press any key
to continue
**Flo-Cal**
Press CLEAR to
change product
Enter new
product #
LIN
Gm 0
No T/C
GO PROVE
LIN
Gm 0
-198.8c
GO PROVE
PRODUCT SELECTOR
0 = LOX
1 = LIN
2 = LAR
3 = LCO2
4 = N20
5 = CCO2
6 = MAPP
7 = LPG