Luminex xPONENT 4.2 for FLEXMAP 3D User Manual

Page 10

Advertising
background image

7. Type a name for the plate in the Plate Name box, or select a saved plate from the Plate

Name list.

NOTE: If a saved plate is selected the results of the new calibration will

override the previous calibration.

8. In the Plate section, click D6 for the 96-well plate. Click H12 for the 384-well plate.

9. In the Reservoir section, click RB1.

10. In the Strip-Wells section, click SD1.

11. Click Auto Adjust Height. The probe automatically adjusts itself and saves it to the plate

you selected.

NOTE: The probe height is automatically set to 0.49 mm. The probe

automatically adjusts this distance from the bottom of the plate or
calibration disks.

NOTE: When you adjust and save the probe height settings for all three

under a plate name, all areas retain the adjustment.

WARNING:

Correct sample probe height is critical to successful sample
acquisition and calibration. Problems with the sample probe can
lead to fluid leaks and inhibit sample acquisition.

CAUTION:

Ensure that the probe height is set correctly before calibrating
the system.

xPONENT

for FLEXMAP 3D

®

6