Upgrade instructions, Install new pc – Luminex LDS 1.7 SP2 to xPONENT 3.1 Rev 2 Upgrade Instructions User Manual
Page 8

Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use
4
These hardware items are part of the upgrade kit only if you purchase
them from Luminex:
•
PC
•
A touchscreen monitor or a flatscreen monitor
•
XYP
These license packages are part of the upgrade kit only if you
purchase them from Luminex:
•
21 CFR Part 11 package
•
LIS package
•
Automation package
•
Security package
•
Extra seats package
Upgrade
Instructions
You must run Luminex xPONENT 3.1 SP1 on a PC that meets the
minimum specifications listed in "System Requirements" on page 3.
To upgrade your PC, contact Luminex Technical Support. See
"Luminex Contact Information" on page 8.
•
If you are installing a new PC, continue with "Install New PC"
below.
•
To upgrade an existing PC that meets the minimum
specifications, continue with "Archive “My Sessions” Folder" on
page 5.
Install New PC
To install the new PC:
1. Close all applications. Perform a complete system shutdown.
Turn the power off and unplug the Luminex analyzer, Luminex
XYP, and PC.
2. Remove all connections from the old PC. Remove the old PC
and set it aside. Set the new PC in place.
3. Connect the keyboard and mouse to an available USB port, and
bar code reader to an available USB or PS2 port.
4. Connect the Luminex XYP serial cable to the back of the XYP
and the computer serial port.
5. Connect the serial end of the Luminex 100/200 serial-to-USB
cable to P1 of the Luminex 100/200 and the USB end to an
available USB port on the computer.
6. Connect the monitor cable to the computer. Touchscreen
monitors will have two cables.
Note: If you did not get a new
PC with your xPONENT 3.1
SP1 upgrade, skip this step.