Output Data Screen
The Output Data tab provides access to 
the most commonly adjusted parameters 
for the transmitter.
Linearization On/Off
This feature corrects for mechanical 
non-linearity of the meter to produce an 
output signal that is precisely proportional 
to flow. Typically the only reason to 
turn off linearization is to collect the 
calibration data with which to populate 
the linearization table on the factory data 
tab. Note that on analog transmitters this 
feature cannot be disabled.
Pulse Output
With linearization enabled the meter’s 
output can be set to a precise Output 
K-Factor where each pulse represents 
a fixed volume. This allows the user to 
take several meters and set them all to an 
exact scaling. With Linearization off, you 
are able to set a fixed number of pulses 
per revolution of the meter. 
Note for quadrature output transmitters, 
the Output K-Factor shown is the output 
on one of the channels (phase A OR phase 
B) however the meter’s total resolution is 
the total number of pulses output on both 
channels (phase A PLUS phase B).