Meriam M203 Altimeter Indicated Air Speed Tester User Manual

Page 27

Advertising
background image

24

11. Press the PRGM key at
YES to save the Re-
Calibration data or at NO to
exit without saving.

YES to save the Re-
Calibration data or at NO to
exit without saving.

12. Press the left arrow key.

Returns to Measure Mode.

RE-CALIBRATION – Restore Factory Defaults

To restore the re-calibration data to the factory defaults, follow these steps:

Keystroke

Display

1. With unit OFF,

press and

hold the MIN/MAX key,

turn the unit on using the

OFF key, then release.

Top line reads “RE-CAL”.
Bottom line reads “EDIT”.

2. Press the up arrow

key twice

Top line reads “RE-CAL”.
Bottom line reads
“RESTORE DEFAULTS”.

3. Press the PRGM

key.

Top line reads
“RESTORE DEFAULTS”.
Bottom reads “YES” or
“NO”.

4. Use the up and

down arrow

keys to select YES or NO

when asked to restore

defaults.

Top line reads
“RESTORE DEFAULTS”.
Bottom reads “YES” or
“NO”.

5. Press the PRGM

key.

Top line reads “RE-CAL”.
Bottom line reads
“RESTORE DEFAULTS”.
Factory defaults have been
restored.

6. Press the left arrow

key.

Returns to Measure Mode.