Health physics division, Passive measurement components ** selectable units, Alarm levels typical one year lifetime with – Mirion Technologies SOR-T User Manual

Page 2: Standard battery user selectable display modes, Backlighted display (option), Periodic exhaustive self-testing, Real time teledosimetry transmission, Flash gamma dose measurement**: 5 cgy to 10 gy, Flash neutron dose management, 1μgy to 10 gy gamma dose rate measurement range

Advertising
background image

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

Mirion Technologies (MGPI) SA

BP 1

F-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 468 4600

F +358 2 468 4601

Mirion Technolgies (RADOS) GmbH

Ruhrstrasse 49

DE-22761 Hamburg

Germany

T +49 (0) 40 851 93-0

F +49 (0)40 851 93 256

Health Physics
Division

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

SOR/T - SOR/R

Active Dosimetry - Dosimeters

www.mirion.com

144258EN-B

FUNCTIONAL CHARACTERISTICS

Redundant architecture with

●

passive measurement components **

Selectable units:

●

cGy; cGy/h; mSv; mSv/h; mrem;

mrem/h

4 confi gurable dose and dose rate

●

alarm levels

Typical one year lifetime with

●

standard battery

User selectable display modes

●

Backlighted display (option)

●

Periodic exhaustive self-testing

●

including the detector

Historical record of measurements

●

and events (750 steps; 10 s;

1 min ;10 min; 1 h; 24 h)

Data storage in EEPROM (qualifi ed

●

> 10 years without battery)

Battery low (16 h) pre-alarm and

●

alarm if the battery is removed

Neck lanyard or clip

●

Real time teledosimetry transmission

●

(up to 1000 m - 3281 ft range) for SOR/R

Training mode included***

●

PHYSICAL CHARATERISTICS

Hp(10) dose equivalent measurements

●

Flash gamma dose measurement**: 5 cGy to 10 Gy

●

Flash neutron dose management**

●

Relative error of fl ash measurement**: ± 30 % over

●

measurement range

Ambient gamma dose measurement range:

●

1μGy to 10 Gy

Gamma dose rate measurement range:

●

from 0.1 μGy/h to 10 Gy/h

Gamma dose rate display:

●

from 1 or 10 μGy/h to 10 Gy/h

Saturation indication (above 10 Gy/h)

●

Relative error of ambient measurement:

●

<± 20 % over the dose measurement range

○

Energy response:

●

<± 20 % in the range 60 keV to 2 MeV

○

<± 50 % in the range 2 MeV to 6 MeV

○

Accredited factory calibration to IEC 17025

●

Accuracy <± 10 % (

●

137

Cs, ~ 25 mSv/h including ± 5 %

of extended uncertainty K=2)

MECHANICAL CHARACTERISTICS

Dimensions: 80.4 x 48 x 9 mm (fl at housing)

●

(3.16 x 1.85 x 0.35 in)

Weight: 55 g (1.94 oz)

●

ENVIRONMENTAL CHARACTERISTICS

-20° to + 50°C (-4°F + 122°F) (normal operating range,

●

standard battery 3V LiMnO

2

CR2450)

-40° to + 50°C (-40°F + 122°F) (option with battery

●

module 3.6V LiSoCl

2

)

TREE protection (SOR/R and SOR/T)

●

Resistant to EMP, EMC, radars

●

Resistant to water immersion (IP67: 1 m / 39.3 in) ,

●

drops, shocks,vibrations, low pressure,

Initial conditions, NBC environmental conditions

○

Complies with the following standards:

●

Meets MIL-STD-810 and MIL-STD-461 requirements

○

Qualifi ed by most of the NATO military laboratories

○

Complies with IEC 1283, ANSI 42-20 and NATO D104

○

NATO supplier n° F8929

Development status: Series production

* LLR : Low level Radiation

** special characteristics of the SOR/T

***a specifi c confi guration software is

required for the XOM reader

SOR inside the
arm-band pouch
(accessory)

SOR version to be
clipped on a pocket

SOR worn round
the neck

Advertising
This manual is related to the following products: