Health physics division, Nuclear characteristics, Functional features – Mirion Technologies SpiR-ID User Manual

Page 2: Identpro/sia identification algorithm, Accessories, Options

Advertising
background image

Health Physics

Division

Since norms, specifications and designs are subject to occasional change, please ask for confirmation of the information given in this publication.

Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179

Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506
FIN-20101 Turku
Finland

T +358 2 468 4600
F +358 2 468 4601

Mirion Technologies (RADOS) GmbH

Ruhrstrasse 49
DE-22761 Hamburg
Germany

T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256

Mirion Technologies Shanghai Branch

Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924

SpiR-ID

Detection & Identification - Risks Prevention

NUCLEAR CHARACTERISTICS

●

Detectors

○

SpiR-ID NaI: 3” x 1.5”

○

SpiR-ID LaBr: 1.5” x 1.5” LaBr

○

2 moderated LiI(Eu) detectors

○

GM tube for high gamma dose rate

●

Energy range

○

25 keV to 3 MeV (gamma)

○

0,025 eV to 15MeV (neutrons)

●

Measurement range

○

<0.01 µSv/h to 10 mSv/h ( 1µR/hr to
1 R/hr)

●

Identification

○

fast digital MCA 1024 channels

○

throughput >100 000 cps

○

range <0.01 to >25 µSv/h (137Cs)

○

detection and ID performance
overpasses ANSI N42-34, IEC62327,
IAEA NSS1 recommendations

●

Continuous spectra acquisition and
stabilization

○

no need of field calibration

●

Continuous count rate analysis
versus dynamic background

●

True dose rate calculation by spectra
weighting

●

Various spectra accumulation mode:

○

sliding / automated / manual

FUNCTIONAL FEATURES

●

Large TFT 3.5” VGA (640 x 480) sun

readable display

○

trend curve screen with level and ID

○

spectra and ID combined screen

○

event list screen with easy scroll

●

Alarm indication

○

sound or voice alarm and chirp

○

vibrator in the handle

○

colored LEDs

○

progressive alarm based on

identification and level

○

easy acknowledge

○

safety alarm

●

Routine mode / Expert mode

●

Embedded GPS

●

Data storage: SD card

●

Data retrieve and remote operation

○

USB link for PC host

●

External alpha/beta probe connection

IDENTPRO/SIA IDENTIFICATION

ALGORITHM

●

Designed for reliable decision with

short spectra

●

Solves mixed isotopes (up to 8)

●

Processes masking scenarios

●

Tolerant to shielding

●

Category, isotope, confidence level

indication

●

ENVIRONMENTAL, ELECTRICAL,
MECHANICAL CHARACTERISTICS

●

Temperature range: –20°C to 50°C

(-4°F to 122°F); -40°C on request

●

Humidity: up to 100%

●

Shock, vibration and drop resistant

●

MIL461D EMI compliant

●

Water and dust ingress IP65

●

Electrical:

○

Li-ion rechargeable, built-in charger

○

autonomy: typical 10 hours

○

accepts primary batteries for

immediate availability

●

Dimensions: 320 X 145 X 175 mm

3

(12.59 x 5.70 x 6.88 cu in)

●

Weight: NaI version 4.9 kg (10.8 lb),

LaBr version 4.02 kg (8.86 lb)

ACCESSORIES

●

Carrying case, carrying strap

●

External main power supply

●

SPIR-Companion software

OPTIONS

●

Alpha – beta probe

●

Embedded GSM / GPRS / Edge / 3G

●

Desk and wall holder

www.mirion.com

145203EN-D

SPIR Companion software

Common to all libraries

Source search library

Nuclear accident library (*)

(*) Firmware version V1.2.3 and up

40K, 226Ra + daughters, 232Th + daughters

NORM

SNM

merged

SNM

expert
Other

Medical

Industrial

Nuclear

U, Pu, 237Np
U, LEU, HEU, 233U, HBPu, MBPu, LBPu,
238Pu, 241Pu, 237Np
Neutron, Bremsstrahlung (ie Beta), Unknown

18F,/511keV, 51Cr, 67Ga, 99mTc, 99Mo,
103Pd, 111In, 123I, 125I, 131I, 133Xe,
153Sm, 201Tl
22Na, 54Mn, 57Co, 60Co, 75Se, 133Ba,
137Cs, 152Eu, 154Eu, 166Ho, 207Bi, 228Th,
241Am

18F/511keV, 60Co, 106Rh, 131l, 132l, 133l,
Xe133, 134Cs, 137Cs, 152Eu, 154Eu

Advertising