Health physics division, Nuclear characteristics, Detectors – Mirion Technologies HDS-101G_GN User Manual

Page 2: Silicon diode for high gamma dose rate measurement, Lii (eu) scintillator for neutron measurement, Energy range, 30 kev to 3 mev (gamma and x-rays), Measurement and identification range (gamma), Extended gamma range, Three user’s profiles: routine, expert, custom

Advertising
background image

Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179

Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506
FIN-20101 Turku
Finland

T +358 2 468 4600
F +358 2 468 4601

Mirion Technologies (RADOS) GmbH

Ruhrstrasse 49
DE-22761 Hamburg
Germany

T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256

Mirion Technologies Shanghai Branch

Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924

HDS-101G/GN

Detection and Identification - Risks Prevention

www.mirion.com

145298EN-D

Health Physics

Division

Since norms, specifications and designs are subject to occasional change, please ask for confirmation of the information given in this publication.

NUCLEAR CHARACTERISTICS

●

Detectors:

○

CsI(Tl) scintillator for spectrometry
and low gamma dose rate measurement

○

Silicon Diode for high gamma
dose rate measurement

○

LiI (Eu) scintillator for neutron
measurement

●

Energy range:

○

30 keV to 3 MeV (gamma and X-rays)

○

0.025 eV to 15 MeV (neutrons)

●

Measurement and identification range (gamma):

○

0.01 µSv/h to 100 µSv/h (1 µrem/h to 10 mrem/h)

●

Extended gamma range:

○

Continuous: 0.1 mSv/h to 10 mSv/h, 10 mrem/h to 1
rem/h,

○

Temporarily: up to 1 Sv/h (100 rem/h)

ALGORITHM PROCESSING

●

Continuous spectra acquisition (0.2s time slot) and
stabilization without source

●

Continuous dose rate and count rate comparison to
background

●

Varying Background Suppression algorithm (VBS)

continuously analyses the spectra shape and rejects
alerts due to sudden background changes

●

NORM Medical Discrimination algorithm (NMD)

categorizes and identifies up to 4 isotopes simultaneously

NORM:

40K, 226Ra and daughters, 232Th and daughters

Medical:

18F, 51Cr, 67Ga, 75Se, 99mTc, 111In, 123I,
125I, 131I, 133Xe, 201Tl

Industrial:

22Na, 57Co, 60Co, 133Ba, 134Cs, 137Cs,
152Eu, 192Ir, 241Am, Bremstrahlung

SNM:

233U, 235U, 237Np, 238U, 239Pu

●

Cumulated spectra can be triggered by alert detection or
on request

FUNCTIONAL FEATURES

●

Three operation modes: detection, search and integration

●

Three user’s profiles: routine, expert, custom

●

Alert threshold configurable for dose rate and/or number
of sigma level increase

●

Radiation safety alarm

●

LCD display with backlight

●

Pre-loaded languages

●

Memory to store more than 1000 events/ measurements
and over 100 spectra in different modes

●

Real time data and spectrum transmission

●

Bluetooth®, RS232 and USB interfaces

●

Audio output for earphone

●

Designed to meet/exceed ANSI 42-33, Type I & II and to
IAEA standards

ELECTRICAL & MECHANICAL CHARACTERISTICS

●

Power supply: 6 x AA batteries (Ni-MH rechargeable),
built-in charger

●

Battery life time: 24 hours (Bluetooth disactivated)

●

Dimensions: 280 mm x 78 mm (11 in x 3.1 in)

●

Weight: 1500 g (52.9 oz)

ENVIRONMENTAL CHARACTERISTICS

●

Temperature range: -20°C to 50°C (-4°F to 122°F)

●

Humidity: < 90% at 42°C (108°F)

●

EMI, shock, vibration and drop resistant

●

Water protection (IP54)

●

CE compliant (RF susceptibility 3 V/m)

ACCESSORIES

●

Shoulder strap

●

USB cable

●

Transportation case

●

HDSMASS and SMI
software

●

Power supply/charger/
wall or car cradle

HDSMass software spectra