Mirion Technologies SGLM 201K User Manual

Radiation monitoring systems, Sglm 201k, Steam generator leak rate monitor

Advertising
background image

Featuring:

radiation monitoring

systems

A Mirion Technologies Division

SGLM 201K

Special Monitors

SGLM 201K

Steam Generator Leak Rate Monitor

The SGLM 201K forms part of the RAMSYS product

line. It has been developed for steam generator leak

rate monitoring. The

16

N count rate is measured by a

scintillation detector (NaI).

The detector is installed in close proximity to the main

steam line and it is thermally insulated to protect it

from sharp temperature transients. The conversion

from

16

N count rate to volumetric activity is done

utilizing coeffi cient established through Monte Carlo

analysis.

The conversion from

16

N volumetric activity to leak

rate (in GPD) is calculated using four correlations. The

correlations are derived by thermohydraulic transport

analysis based upon data provided by the user.

APPLICATIONS

• Barrier leak control

○

Secondary circuit monitoring

RELATED MONITORS

• SGLM 202K: version with lead shield for WWER

nuclear power plants

FEATURES

●

16

N monitoring when reactor power exceeds 25%

●

Gross gamma energy monitoring when reactor

power < 25%

●

Spectrum stabilization against temperature and aging

drifts

●

16 confi gurable windows over the measurement range

●

1024 channels analysis

●

Available with or without display and local signaling

●

Seismically qualifi ed

●

More than 500 SGLM channels in operation worldwide

●

Defi ned for mild environment conditions