Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion Technologies (MGPI) SA

Route d’Eyguières

FR-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna, GA 30082

USA

T +1 770 432 2744

F +1 770 432 9179

Mirion Technologies (MGPI H&B) GmbH

Landsberger Strasse 328a

DE-80687 Munich

Germany

T +49 (0) 89515 13-0

F +49 (0) 89515 13 169

Mirion Commercial (Beijing) Co., Ltd.

Shanghai Jiangchang Commercial Branch

Room 801, 78 Jiangchang SanLu

Zhabei District, Shanghai 200436

PR of China

T +86 21 6180 6920

F +86 21 6180 6924

SAM 202K

Spectrum Analysis Monitors

PHYSICAL CHARACTERISTICS

●

Radiation detected: gamma

●

Detector: 3”x2” NaI(TI) scintillator + PMT

●

Energy range: 100 keV to 7 MeV

●

Spectrometry analysis: 1024 channels

●

Typical measurement range: 6 decades between
3.7 10

+2

to 3.7 10

+12

Bq/m

3

(10

-8

to 10

+2

μCi/cc)

ENVIRONMENTAL CHARACTERISTICS

●

Long term temperature: +10°C to +40°C
(+50°F to +104°F)

●

Periodic temperature: -5°C to +55°C

(+23°F to +131°F)

●

MTBF: > 20 000 hours

●

TID (processing unit): 100 Gy (10

+4

rad)

●

TID (detector): 10

+3

Gy (10

+5

rad)

●

Protection index: IP65 and IK07

MECHANICAL CHARACTERISTICS

●

Dimensions:

○

Processing unit: 390 mm x 196 mm x 187 mm
(15.3 in x 7.7 in x 7.3 in)

○

Detector: 329 mm x 388 mm x 513 mm
(13 in x 15.3 in x 20.2 in)

●

Weight:

○

Processing unit: 8 kg (17.6 lb)

○

Detector: 220 kg (485 lb)

●

Color: gray RAL 7030 (decontaminable paint)

ELECTRICAL CHARACTERISTICS

●

Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz

●

Data link outputs: 1 RS232 and 2 isolated RS485

●

Alarm relays: 3 SPDT relays

●

I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)

SIGNALING (Applicable to LPDU only)

●

Alphanumeric display: measurement, status…

●

Sound alarm: buzzer 90 dBA at 1 meter

●

Visual alarm: 3 lights (red, yellow, green)

REFERENCE STANDARDS

●

Environmental: IEC60780, IEEE323

●

Seismic: IEC60980, IEEE344

●

EMC: 2006/95/CE, 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4

VERSIONS

●

230 Vac or 120 Vac

●

LPDU or LPU

●

Without protection, 2 π/5 cm (2 π/2 in) or 4 π/5 cm
(4 π/2 in) detector lead shield

●

With or without RS485 junction box

●

Detector cable length: from 20 m (65.6 ft)
to 100 m (328 ft)

●

Junction box cable length: 2 m (6.56 ft), 5 m (16.4 ft)
or 10 m (32.8 ft)

ACCESSORIES

●

Calibration tools

●

Software

●

USB converters

●

Seismic qualifi ed wall mounting bracket

●

Detection sub-assembly support

www.mirion.com

144425EN-E