Physical characteristics, Radiation detected: gamma, Detector: 1¼”x1” nai(ti) scintillator + pmt – Mirion Technologies IM 201S User Manual

Page 2: Am source, Energy range: 100 kev to 3 mev, I, eγ 364 kev), 1024 channels spectrum, Bq/m, Μci/cc) environmental characteristics, Rad) pneumatic characteristics

Advertising
background image

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion Technologies (MGPI) SA

Route d’Eyguières

FR-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna, GA 30082

USA

T +1 770 432 2744

F +1 770 432 9179

Mirion Technologies (MGPI H&B) GmbH

Landsberger Strasse 328a

DE-80687 Munich

Germany

T +49 (0) 89515 13-0

F +49 (0) 89515 13 169

Mirion Commercial (Beijing) Co., Ltd.

Shanghai Jiangchang Commercial Branch

Room 801, 78 Jiangchang SanLu

Zhabei District, Shanghai 200436

PR of China

T +86 21 6180 6920

F +86 21 6180 6924

IM 201S

Particulate, Gas & Iodine Monitors

PHYSICAL CHARACTERISTICS

●

Radiation detected: gamma

●

Detector: 1¼”x1” NaI(TI) scintillator + PMT +

241

Am

source

●

Iodine cartridge: 57.7 mm (2.27 in)

●

Energy range: 100 keV to 3 MeV

●

Typical energy windows: 314 - 414 keV
(

131

I, Eγ 364 keV)

●

1024 channels spectrum

●

Typical measurement range: 3.7 to 3.7 10

+6

Bq/m

3

(10

-10

to 10

-4

µCi/cc)

ENVIRONMENTAL CHARACTERISTICS

●

Long term temperature:

+10°C to +40°C (+50°F to +104°F)

●

Periodic temperature:
-5°C to +55°C (+23°F to +131°F)

●

MTBF: > 20 000 hours, with preventive maintenance

●

TID: 100 Gy (10

+4

rad)

PNEUMATIC CHARACTERISTICS

●

Standard fl ow rate: 35 l/min (1.24 scfm)

●

Pressure drop: according to the fi lter dust loading

MECHANICAL CHARACTERISTICS

●

Dimensions: 1280 mm x 830 mm x 680 mm (50.4 in x
32.7 in x 26.8 in)

●

Weight: ~ 300 kg (~ 661 lb)

●

Color: gray RAL 7030 (decontaminable paint)

●

Inlet tube connection: Ø 12 mm OD (1/2 in)

●

Outlet tube connection: Ø 12 mm OD (1/2 in)

ELECTRICAL CHARACTERISTICS

●

Power supply: refer to possible versions

●

Data link outputs: 1 RS232 and 2 isolated RS485

●

Alarm relays: 3 SPDT relays

●

I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)

SIGNALING

●

Alphanumeric display: measurement, status…

●

Sound alarm: buzzer 90 dBA at 1 meter

●

Visual alarm: 3 lights (red, yellow, green)

REFERENCE STANDARDS

●

Nuclear: IEC60761, IEC61171

●

Environmental: IEC60780, IEEE323

●

Seismic: IEC60980, IEEE344

●

EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4

VERSIONS

●

230 Vac or 230 Vac + 400 Vac 3Ø or 120 Vac
+ 400 Vac 3Ø

●

Solenoid check source

●

With or without dust fi lter holder

●

Gas grab sampler ports

ACCESSORIES

●

Calibration tools

●

Software

●

USB converters

●

Ethernet

●

ESD board

www.mirion.com

144106EN-D