Multichannel Systems MEA1060-Up Manual User Manual

Page 25

Advertising
background image

Appendix

21

7.1.2 MEA1060-Up

Operating temperature

10 °C to 50 °C

Storage temperature

0 °C to 50 °C

Dimensions (W x D x H)

165 mm x 165 mm x 14.5 mm (19 mm with sockets)

Weight

600 g

Supply voltage (external power supply)

± 6 VDC to ± 9 VDC

Supply current

max. 220 mA, typically ± 150 mA

Number of stimulus input channels

60

Maximum recommended stimulus input
voltage

± 5 V

Number of analog input channels

60

Input voltage

-4 mV to 4 mV (with respect to the standard gain of
1200)

Input impedance

10

11

 parallel to 10 pF

Input noise

< 800 nV

RMS

Noise density

e

n

=

Number of analog output channels

60

Output voltage

± 5 V

Output current

10 mA

Output impedance

300

Bandwidth

1 Hz to 3000 Hz (standard), 1 Hz to 5000 Hz (custom)

Gain

1200 (standard), 500 to 5000 (custom)

Microscope compatibility

Upright microscopes

Distance focal plane to microscope table

8 mm

7.1.3 Heating

Element

The heating element is mounted onto the bottom side of the base plate of the MEA1060
amplifier. It is used for keeping the temperature in the MEA culture chamber stable during
recordings.

Input voltage

max 10 V

Input current

max 2 A

Temperature sensor

PT 100

Resistance

15 ± 2

Heating temperature

Ambient temperature (min 10 °C) to 50 °C

Accuracy

0.1 °C (in the range of 20 °C to 40 °C)

Recovery time

0.5 min to 2 min

Calibration constant

T = c * (R R

0

) / R

0

for c = 259.7 °C

Thermal resistance

6 °C / Watt

Typical time constant

50 s