Roboinject Manual

32

Air Pressure Section

Check the air pressure. It should be set to a minimum value of 3000 hPA,
and should never exceed a value of 4 bar.

Injector Mount = z-Axis Section

Click "Reference". The z-axis will perform its reference movement and will stop in home position.
The depicted Z value is 0 μm.

Click "Home" to move the z-axis into home position. The home position for the z-axis is defined
as the position when the injection device is at its topmost position. The depicted Z value is 0 μm.

Click "Tube". The z-axis moves to a position from where a sample can be taken up from one of
the reservoir tubes.

Warning: Never try to move the z-axis to any tube position before moving the carrier
to the respective tube. Otherwise there is the risk of severe damage of the injector.

Click "Impalement". Select the "Depth" of impalement from the drop down menu.

Select "Rinse 1", "Rinse 2" or "Waste". The actual position of the z-axis will be displayed
in "Z = x μm".

Use the arrow buttons to move the z-axis stepwise in 20 μm "Slow" or 200 μm "Fast" steps
up and down. The "Z" value shows the actual position of the z-axis in μm relative to the home
position (Z = 0).