7.0 CALIBRATION

• Calibration is recommended for best accuracy. It is suggested that the

scale be calibrated when first installed and any time the scale is moved
to a different environment.

• To ensure best accuracy calibrate before use.
• Ensure weights are ready before calibrating, the weight should be of a

known accuracy, for example OIML class M1 or Class 4 of ASTM E617.

• Do not calibrate the scale if no calibration weight is available.
• Calibration can only be accessed if the Calibration function has been

enabled as described in section 6.0.

7.1

AUTOMATIC

EXTERNAL

CALIBRATION

• In the weighing mode, press [Func] key to access calibration mode. The

display will show “CAL”.

• Press [Zero] key, an access code (Pin code) is required to continue, The

display will show “Pin” .

• To enter the access code press the [Func] key, [Unit] key and[Tare] key

in sequence, then press the [Zero] key to confirm the pin code.

• The display will show the weighing unit last used to calibrate the scale,

“S-Unit kg”(or ”S-Unit Lb”) ,press [Unit] key to change to lb’s or kg’s for
the calibration weighing unit value.

• Press [Zero] key to start the zero calibration. The display will show “CAL

0” and when the stable symbol is displayed press [Zero] key to confirm.

• The display will now show the calibration mass expected. ”CAL XX Kg “

or ”CAL XX Lb “

.Press [Tare] key to change the calibration values if

necessary. Each model has 3 different kg and lb calibration selection

values.