Specifications, Error messages – Nova-Tech pH-CON 300 User Manual

Page 23

Advertising
background image

10. Specifications

Mode

pH

Temperature

Conductivity

TDS

0 to 19.99 µS

0.00 to 9.99 ppm

–2.00 to

0 to 100°C /

0 to 199.9 µS

10.0 to 99.9 ppm

Range

16.00 pH

32.0 to 212.0°F

0 to 1999 µS

100 to 999 ppm

0 to 19.99 mS

1.00 to 9.99 ppt

0 to 199.9 mS

10.0 to 199.9 ppt

0.01 µS, 0.1 µS, 0.01 ppm, 0.1 ppm,

Resolution

0.01 pH

0.1°C or °F

1 µS, 0.01 mS,

1ppm, 0.01 ppt,

0.1 mS

0.1 ppt

Accuracy

±0.01 pH

±0.5°C / ±0.9°F

±1% full scale

±1% full scale

up to 5 points

offset in 0.1°

up to 5 points

up to 5 points

Calibration

(pH 1.68, 4.01,

increments

(one point

(one point

7.00, 10.01, 12.45)

up to 5°

per range)

per range)

pH slope: 80 to 110%

Conductivity cell constant: K = 1.0

Conductivity/TDS temperature coefficient: adjustable from 0.0 to 10.0% per °C

Conductivity/TDS normalization temperature:
adjustable from 15 to 30°C / 59 to 86°F

Conductivity-to-TDS conversion factor: adjustable from 0.4 to 1.0

Temperature compensation: automatic or manual from 0 to 100°C / 32 to 212°F

Display: dual LCD

Operating temperature: 0 to 50°C / 32 to 122°F

Power: four 1.5 V AAA batteries (included)

Battery life: > 200 hours continuous use

Probe connectors:
pH/mV: BNC connector
Conductivity/TDS/Temperature: notched six-pin connector

Dimensions:
Meter: 7.5"L x 3.5"W x 1.75"H

(19.1 cm x 8.9 cm x 4.5 cm)

Boxed: 9.2"L x 9.2"W x 2.75"H

(23 cm x 23 cm x 7 cm)

Probe: 6.8"L x 1.3" dia (17.3 cm L x 3.2 cm dia)

Weight:
Meter: 1.0 lb (0.5 kg)
Boxed: 2.0 lbs (0.9 kg)

45

44

9. Error Messages

LCD

Indicates

Cause

Solution

Display

Err

Unrecognized input

Wrong input in

Release key. Select

annunciator

from keypad

selected mode.

valid operations
depending on mode.

CAL & Err

Calibration error

Wrong value input

Check your input value,

annunciators

at calibration.

clean probe.

blink

Dirty probe.

See Calibration sections or
Probe Maintenance section.

Battery

Low battery level

Need new batteries

Clean battery contacts.

indicator blinks

or battery

Replace batteries with fresh

connection is bad.

ones, noting polarity.

If error persists, or the meter shows incorrect values, return the meter.
See “Warranty” and “Return of Items” on page 52.

For a complete diagram of the display, see page 4.