1. Introduction:

The Autofil

TM

filtration system has been designed for use in laboratory

environments. The materials of construction are compatible with common
disinfectants and sanitizing agents. The smooth contoured design is readily
cleaned by spraying or applying a disinfectant and wiping as required.


Under normal operating conditions, fluids should not come in contact with the

internal plumbing of the Autofil

TM

pedestal or cradle ring stand. In the event fluids

do come in contact with the internal plumbing, the following procedure can be
followed to decontaminate the system.

2. Preparing the Sanitizing Agent:

a) The following sanitizing agents can be used:

Iodophors (follow manufacturers recommended dilution)

Alcohols (effective dilution is 70-85%)

Commercially available disinfectants

NOTE:
Use of hypochlorite (bleach) is not recommended. Prolonged exposure
to hypochlorite may cause corrosion to internal stainless steel
components. Adequate water rinsing is necessary if hypochlorite
solution is used.

b) Prepare 500mL of sanitizing agent in a proper flask or container following

manufacturer’s recommendation.

3. Prepare the pedestal or cradle ring stand for sanitation (refer to illustration below)

a) 1) Inserting a blunt point into the hole and gently remove rubber gasket

from the pedestal or cradle ring stand. 2) Insert vacuum tube adaptor into
main vacuum port.