Flowmeter kits, Disassembly and assembly – Nova-Tech RITEFLOW FLOWMETERS AND KITS User Manual

Page 10

Advertising
background image

9

The Aluminum model has

1

/

8

” polypropylene hose barb fittings which accept input/output tubing

from

5

/

16

” to

3

/

8

” i.d. The Teflon

®

model has

3

/

8

” o.d. straight glass connectors for input/output

tubing. Input should be connected to the bottom; output tubing to the top.

Mounted Flowmeters are available in Teflon

®

or Aluminum models in both 65mm and 150mm

sizes.

FLOWMETER KITS

Kits are designed to allow the use of the entire Riteflow

®

line of 150mm Flowmeters on one

stand for a wide range of flow measuring capability.

Mounted Aluminum 150mm Flowmeter Kits (H40409-0000) include an aluminum frame, an acrylic
leveling base, large orifice for high flowrate measurements, 4 flowtubes (Size No. 1,2,3 and 5), 8
floats (2 in each of the flowtubes), top and bottom Viton gaskets for all Flowmeters, two support
brackets, two mounting rods, two rod clamps, 5 vials for float storage and a Flowmeter cleaning
kit (94040-9026).

Mounted Teflon

®

150mm Flowmeter Kits (H40408-0000) include a Teflon frame, an acrylic leveling

base, 4 flowtubes (Size No. 1,2,3 and 4), 9 floats (2 in each of the flowtubes and

1

/

4

” dia. Carboloy

in a vial), two support brackets, two mounting rods, two rod clamps, 5 vials for float storage and a
Flowmeter cleaning kit (94040-9026).

DISASSEMBLY AND ASSEMBLY

Each Guarded and Mounted Flowmeter contains a flowtube, which can be removed or replaced
using the following procedure (See Figure 1).

GUARDED FLOWMETER:
1. Unscrew two end bushings.

2. Remove flowtube from plastic guard.

3. Reassemble in the following order:

a. Insert top rubber gasket (P) in upper bushing (M).

b. Screw plastic tube (N) on upper bushing.

c. Insert flowtube (B) into plastic tube until top of Flowmeter seats against rubber gasket.

d. Insert rubber gasket (P) into lower bushing (M).

e. Screw lower bushing (M) into plastic tube until lower end of the Flowmeter seats against

rubber gaskets.

f. Plastic tube will almost completely cover the thread on bushings and provide a tight seal

against the inner rubber gasket.

Advertising