Nova-Tech pH Tutor Benchtop Meter User Manual

Page 2

Advertising
background image

5.

Repeat with other buffers if necessary. Rinse electrode in tap water before dipping into
next buffer.

Note: The calibration mode allows you to perform up to three calibration points before returning
to the measurement mode automatically. However, if you opted to have only one or two calibration
points, simply skip the remaining calibration points by exiting to the measurement mode by
pressing the CAL / MEAS key.

pH Measurement

Automatic Temperature Compensation (ATC) (Optional Probe Purchase)

For ATC measurements, simply attach a temperature probe into the meter.
The “ATC” annunciator lights up on the LCD. Submerse the probe in the liquid
you are measuring so that the sample temperature can be recorded and
compensated for.

Manual Temperature Compensation (MTC)

For MTC, you must disconnect the temperature probe.

1.

While pressing the ENTER key, switch on the meter by pressing the ON/OFF key.

2.

Release the ENTER key. You will enter the setup mode with the display showing “CAL”

3.

Press ENTER key to enter the Temperature compensation mode. The upper display act as
the adjustable temperature setting and the lower display shows the default temperature
value of 25°C or indicates the last set temperature setting.

4.

Check the temperature of your sample using an accurate thermometer.

5.

Press

or

key to offset the temperature to the measured value from step 4.

6.

Press the ENTER key to confirm the setting and to return to the measurement mode. The
meter will compensate pH reading for the manually set temperature.

Note: To abort MTC calibration, do not press the ENTER key in step 6. Press the ‘CAL / MEAS’
key instead.

Taking Measurements

Be sure to remove any electrode soaker bottle or protective rubber cap from the electrode
before measurement.

1.

Press ON/OFF key to switch on meter. The MEAS annunciator appears on the top of the
LCD. When used with temperature probe, an ATC annunciator will appear in the lower right-
hand corner to indicate Automatic Temperature Compensation.

2.

Dip the electrode about 2 to 3 cm into the test solution. Stir and let the reading stabilize.

3.

“READY” annunciator will be displayed once the reading stabilizes. Note the reading on the
display.

User Reset

You can reset the pH calibration to the factory default by using the user reset function. Buffer
set selection and user temperature calibration are not affected by the user reset function.

1.

Switch off the meter.

2.

While pressing the ENTER key, switch on the meter by pressing the ON/OFF key.

3.

Release the ENTER key. You will enter the setup mode with the display showing “CAL”

4.

Press the

or

key to scroll through the setup page until you view the “rSt” display.

5.

Press the ENTER key to enter the user reset option page.

6.

Use the

or

key to decide your option on the user reset function.

nO deactivates reset function

YES activates the reset function

7.

Press the ENTER key to confirm your option.

8.

If you have selected ‘YES’, the display will flash momentarily and proceed to the measurement
mode with the calibration reset back to factory default value.

9.

If ‘nO’ is selected, the unit will proceed to the measurement mode without any user reset
performed.

Viewing Electrode Properties

Your pHTutor lets you check the electrode parameters for diagnostic purposes. You can view
the offset and slope of the electrode to determine its effectiveness.

1.

While pressing the ENTER key, switch on the meter by pressing the ON/OFF key.

2.

Release the ENTER key. You will enter the setup mode with the display showing “CAL”

3.

Press the

or

key to scroll through the setup page until you view the “ELE” display.

4.

Press the ENTER key to enter the electrode properties viewing page. The display will show
the mV offset value of the electrode.

5.

Press the ENTER key again to view the slope in % of the electrode.

6.

To return to the “ELE” page, press the ENTER key.

7.

Press

or

key to select a new program OR press CAL/MEAS key to return to the

measurement mode.

Self-Diagnostic Messages

Over range / Under

Or / Ur

Electrode is not in contact with solution or

range signal

electrode is failing.

Measured pH value or temperature value exceeds
its specified maximum or minimum value

Error Message

Err

Wrong key selected

Err

pH calibration error of attempting to confirm a
calibration value which is not within the specified
calibration window

Accessories

Replacement meter and accessories

Item Description

Eutech Instruments

Oakton Instruments

Order Code No.

Order Code No.

CyberScan pHTutor Bench pH/

EC-PHTUTOR

35619-30

Temp Meter

Temperature Probe for pHTutor bench

EC-PH5-TEMB01P

35613-05

meter, 1m cable length

Electrode Stand with Swivel Arm &

EC-PH-ELSTDC

35617-50

Base Plate

110/120VAC power adapter, 50/60 Hz

EC-120-ADA

35615-07

220/230VAC power adapter, 50/60 Hz

EC-220-ADA

35615-08

pH Buffers/Sachets, Reference Electrolyte & Others

Item Description

Eutech Instruments

Oakton Instruments

Order Code No.

Order Code No.

pH 4.01 Buffer Solution (480 ml bottle)

EC-BU-4BT

00654-00

pH 7.00 Buffer Solution (480 ml bottle)

EC-BU-7BT

00654-04

pH 10.01 Buffer Solution (480 ml bottle)

EC-BU-10BT

00654-08

pH 4.01 Buffer Sachets

EC-BU-4BS

35653-01

(20 ml x 20 pcs. per box)

pH 7.00 Buffer Sachets

EC-BU-7BS

35653-02

(20 ml x 20 pcs. per box)

pH 10.01 Buffer Sachets

EC-BU-10BS

35653-03

(20 ml x 20 pcs. per box)

Reference Electrolyte (KCl with

EC-RE001

05992-47 (125 ml)

Ag/AgCl) (480 ml bottle)

Reference Electrolyte (4MKCI)

EC-RE002

05992-48 (125 ml)

for Calomel (Hg/ Hg2Cl2) electrodes
& double-junction electrodes

Storage Solution for pH Electrode

EC-RE005

00653-04

(480 ml bottle)

Protein Cleaning Solution (480 ml bottle)

EC-DPC-BT

00653-06

Replacement Electrodes

Item Description

Eutech Instruments

Oakton Instruments

Order Code No.

Order Code No.

General purpose, Epoxy-body

EC-FC72522-01B

35641-51

double junction pH Combination
Electrode, 12 X 110 mm,
1m cable length

General purpose, Epoxy-body

EC-FC73529-01B

35811-71

“3-in-1” pH/Temp. Combination
Electrode, 12 X 110 mm,
1m cable length

Glass body Refillable pH

EC-FG73504-01B

05997-10

Combination Electrode:
Annular ceramic reference junction
with protective sensor guard,
1 m cable length

Warranty

The pHTutor are warranted to be free from manufacturing defects for 3 years and electrode
module for 6 months. If repair, adjustment or replacement is necessary and has not been the
result of abuse or misuse within the time period, please return the product – freight prepaid – and
correction will be made without charge. Out of warranty products will be repaired on a charge
basis.

Return of Items

Authorization must be obtained from your distributor before returning items for any reason.
When applying for authorization, please include information regarding the reason the item(s) are
to be returned.
Note: We reserve the right to make improvements in design, construction and appearance of
products without notice. Prices are subject to change without notice.

Specifications

pHTutor

Specifications

pH Range

0.00 to 14.00 pH

Resolution

0.01 pH

Relative Accuracy

0.01 pH

Calibration Points

Up to 3 points with Auto-buffer recognition

Buffer Set Standard Selection

USA- 4.01/7.00/10.01NIST- 4.01/6.86/9.18

Calibration Window

+/-1.00 pH (pH 4.01 & pH 10.01),+/-1.50 pH (pH 7.0)

(USA Buffer Set Standard)

Calibration Window

+/-1.00 pH (pH 4.01 & pH 9.18),+/-1.25 pH (pH 6.86)

(NIST Buffer Set Standard)

Temperature Compensation

Automatic / Manual (0 to 100 °C)

Temperature Range

0 to 100 °C

Temp Resolution

0.1 °C

Temp Accuracy

+/- 0.3 °C

Temp Calibration Window

+/- 5 °C

pH Slope and Offset Display

Yes

User reset

Yes

Non Volatile Memory Backup

Yes

LCD Display

Custom Dual LCD

(1 x 4.5 digits, 1 x 3.5 digits, annunicators)

Power Requirement

AC/DC 9V Adapter (110 VAC/220 VAC, 50-60Hz)

Dimensions / Weight

Meter : 18 x 23 x 6 cm / 1250 g
Boxed : 40 x 26 x 9 cm / 1950 g