Control Company 4481 INFRARED THERMOMETER WITH WRISTBAND User Manual
Control company
 
 
 
SPECIFICATIONS 
Range: 
–67° to 482°F / –55° to 250°C
Resolution: 0.1° (-9.9° to 199.9°) 1° (otherwise) 
 
OPERATION 
Press and continue to hold down the SCAN button to 
take temperature readings. The unit will take 
temperature readings while the SCAN button is pressed 
and held. 
 
After the SCAN button is released, the last temperature 
measured will be displayed for approximately 20 
seconds, the display will then turn off automatically. 
 
SELECTING UNITS OF MEASURE 
Press and release the SCAN button to turn on the 
display. Once the display is on (and the SCAN button 
has been released), press the MODE button four (4) 
times (°F or °C will flash on the display). Press the 
SCAN button to switch the unit of measure. 
 
LASER TARGETING 
The thermometer will take temperature measurements 
without activating the laser targeting by partially 
pressing and holding the SCAN button. To take 
temperature measurements and activate the laser 
targeting, completely press and hold the SCAN button. 
CAUTION: Eye damage may result from direct 
exposure to laser light. 
 
DISTANCE SPOT/RATIO 
There are other factors that may affect measurement 
accuracy. The target must completely fill the spot 
diameter seen by the infrared sensor; otherwise readings 
will be influenced by the surface surrounding the target. 
The ratio of the distance to the size of the spot being 
measured is 6:1. For example, an object’s diameter of 
100mm can be measured from 600mm. 
 
MIN/MAX MODES 
MIN Mode – While in the MIN mode, the unit will 
display only the minimum temperature that has been 
measured while the SCAN button is being pressed. The 
display is updated as new minimum temperatures are 
measured. 
 
To activate MIN mode: 
1. Press and release the SCAN button to turn on the 
display.
2. Press the MODE button until MIN appears on the
display.
3. Press and hold the SCAN button to begin taking
measurements.
 
MAX Mode – While in the MAX mode, the unit will 
display only the maximum temperature that has been 
measured while the SCAN button is being pressed. The 
display is updated as new maximum temperatures are 
measured. 
 
To activate MAX mode: 
1. Press and release the SCAN button to turn on the 
display.
2. Press the MODE button until MAX appears on the
display.
3. Press and hold the SCAN button to begin taking
measurements.
 
LOCK FUNCTION 
The lock function allows the thermometer to take 
measurements without having to press and hold the 
SCAN button. When the lock function is activated, the 
 
 
 
 
 
 
unit will take continuous temperature measurements for 60 
minutes and will then turn off automatically to preserve 
battery life. 
 
To activate the lock mode: 
1. Press the SCAN button to turn on the display. 
2. Press the MODE button until LOCK flashes on the 
display.
3. Press and release the SCAN button to begin taking
temperature measurements (LOCK will appear solid 
{not flashing}, to indicate that the unit is in the lock 
mode and is taking measurements).  
 
Laser Targeting is not available while in the LOCK mode. 
To exit the lock mode, press and release the SCAN button 
(LOCK will no longer appear on the display). 
 
EMISSIVITY 
The emissivity adjustment is optional. Emissivity 
adjustments are used to provide a truer temperature 
reading. Different materials radiate infrared energy at 
slightly different temperatures. The emissivity adjustment is 
used to compensate for different types of materials. The 
default emissivity of 0.95 will cover 90% of typical 
applications. 
 
The emissivity table provides a guide of different emissivity
values for different materials.
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
When the emissivity of an object is unknown use a non-
infrared thermometer, such as a thermometer with a 
surface probe to measure the object’s surface 
temperature. Adjust the emissivity until the temperature of 
the infrared thermometer matches the temperature of the 
surface probe. The resulting emissivity value from this 
method may be used to measure similar materials. 
 
To adjust the emissivity (optional): 
1. Press and release the SCAN button to turn on the 
display.
2. Press the MODE button five (5) times [95E (or
whatever emissivity value has been set) will appear on 
the display]. 
3. To adjust the emissivity value, partially press the SCAN
button to advance the value. (Completely pressing the 
SCAN button while adjusting the emissivity will activate 
the laser targeting). Once the maximum value of 100 is 
reached, the display will roll over to the minimum value 
of 5. 
 
 
 
 
 
 
 
 
TRACEABLE® INFRARED THERMOMETER WITH WRISTBAND INSTRUCTIONS
The emissivity can be set from 0.05 to 1.00 (5 to 100 on 
the display). 
4. Once the desired emissivity value appears on the
display, press the MODE button to confirm the value.
This emissivity value will be used for all temperature
measurements until the value is reset.
 
DISPLAY MESSAGES 
HI 
Appears when the temperature being measured is 
outside the temperature range of the unit (higher 
limit). 
LO
Appears when the temperature being measured is 
outside the temperature range of the unit (lower 
limit). 
Er2 Appears when the unit is exposed to rapid changes
in ambient temperature.
Er3 Appears when the ambient temperature is below
32°F (0°C) or above 122°F (50°C).
Er
Appears when the unit needs to be reset. To reset 
the unit, remove the batteries, wait 2 minutes and 
then reinstall the batteries. 
 
ALL OPERATIONAL DIFFICULTIES 
If this thermometer does not function properly for any 
reason, replace the batteries with new high-quality 
batteries (see Battery Replacement section). Low battery 
power can occasionally cause any number of “apparent” 
operational difficulties. Replacing the batteries with new 
fresh batteries will solve most difficulties. 
 
BATTERY REPLACEMENT 
Erratic readings, faint readings, no display, or 
appearing on the display are all indications that the batteries 
must be replaced. Open the battery covers by turning them ⅛ 
of a turn in the direction of the OPEN arrows (clockwise). (Be 
careful not to turn the cover more than ⅛ of a turn, as this 
may damage the battery cover.) Remove the exhausted 
batteries and replace them with two (2) new CR2032 lithium 
batteries. Make certain to install the new batteries with the 
positive (+) side facing out. Replace the battery covers by 
lining up the tabs and turning the cover approximately 1/8 of a 
turn counter-clockwise until the cover snaps into position. (Be 
careful not to turn the cover after is has snapped into position, 
as this may damage the battery cover.) 
 
 
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714
Fax 281-482-9448
E-mail [email protected]
www.control3.com