Nova-Tech 4373 TRACEABLE MEMORY_WATERPROOF THERMOMETER User Manual

Control company

Advertising
background image

Traceable

®

is a registered trademark of Control Company ©2007 Control Company. 92-4373-00 Rev. 2 101607

SPECIFICATIONS
Display:

LCD, ½” digits

Range:

–58 to 572°F (–50 to 300°C)

Resolution: 0.1° from –20 to 200°, 1° outside this range
Accuracy:

±1°C between –20 and 100°C

Power:

G-13 silver-oxide battery

Size:

4⅝ x 2⅞ x 1 inches

Accessories
Supplied:

Stainless steel probe, battery, stand, clip, Traceable

®

Certificate, and

instructions.


DESCRIPTION
1.

Display

2.

°C/°F Button

3.

On/Off Button

4.

Memory Button

5.

H/T Button

6.

Light Button

7.

Max/Min Button

8.

Alarm Off Button

9.

Record Button

10. Clear Button
11. Low Alarm Button
12. High Alarm Button

OPERATION
1.

Turn on the unit by pressing the On/Off Button (3, Fig. 1) once.

2.

Press the °F/°C Button to select the desired temperature unit. (2, Fig. 1)

3.

Immerse the probe tip at least 10 millimeters into the desired substance to
obtain temperature reading.

4.

Press the H/T button (5, Fig .1) once to “freeze” the display at the current
temperature reading. “HOLD” flashes in the upper portion of the display
indicating that the unit is in hold mode. Press the H/T button again to release the
reading.

5.

Press the MAX/MIN button (7, Fig. 1) once to recall the minimum temperature.
“MIN” will appear on the display.

6.

Press the MAX/MIN button once again to recall the maximum temperature.
“MAX” will appear on the display.

7.

After 3 seconds, the unit will automatically return to the current temperature and
“Min” or “Max” will no longer appear on the display.

8.

To back light the display, press the LIGHT button (6, Fig. 1). The display will
light for 3 seconds.

9.

The unit will automatically turn itself off after one hour to conserve power. To
turn off this function, press Clear (10, Fig. 1) and On/Off (3, Fig. 1). To restore
this function, turn the unit off and back on again by pressing the On/Off button
two times.

10. A stand is located on the back of the unit. Simply pull the bottom of the stand

away from the unit to extend the stand for use. Press the stand back into the unit
until it snaps to close. Also located on the back of the unit is a clip.

11. With the stand extended, pull it gently away from the unit to expose the battery

compartment and probe cable.

SETTING THE ALARM
1.

Press and hold the Low Alarm Button (11, Fig. 1) until the desired temperature is
reached. “LOW” will appear in the display. When the current temperature is at or
below this set value, an alarm will beep.

2.

The alarm will continue to beep four times every minute until the Alarm Off
Button (8, Fig. 1) is pressed or the current temperature returns to an “in range”
reading.

3.

Press and hold the High Alarm Button (12, Fig. 1) until the desired temperature
is reached. “HIGH” will appear in the display. When the current temperature is at
or above this set value, an alarm will beep.



4.

When both a High and Low alarm temperature are set, “HI LOW” will appear in
the lower portion of the display.

5.

To clear the alarms, press the Alarm Off button (8, Fig. 1).


MEMORY
1.

Press the H/T button (5, Fig. 1) to hold the current temperature. Press the
Memory button (4, Fig. 1) to store the temperature reading into memory. “Mem
1” will appear on the display. Repeat these steps to store additional
temperatures. Up to 10 readings can be stored.

2.

To recall a stored reading, press the Recall Button (9, Fig. 1) once. The
temperature stored in Memory One will be displayed. Each subsequent press of
the Recall Button will show the next stored reading.

3.

To clear all stored memories, press and hold the Clear Button (10, Fig. 1) for 5
seconds.

LOW BATTERY
Erratic readings, a faint display, or no display are all indicators that the battery is low
and needs to be replaced. To replace the battery, flip open the stand on the back of
the unit and pull it out to expose the battery compartment. Use a coin to turn and
remove the battery cover. Insert a new G-13 battery. Incorrectly installed batteries may
damage electronics. Replace the Battery Cover.

ALL OPERATIONAL DIFFICULTIES
If this thermometer does not function properly for any reason, please properly replace
the battery with a new G-13 battery (see Low Battery section, above). Low battery
power can occasionally cause any number of “apparent” operational difficulties.
Replacing the battery with a new fresh battery will solve most difficulties.


















WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:

CONTROL COMPANY

4455 Rex Road

Friendswood, Texas 77546 USA

Ph. 281-482-1714 Fax 281-482-9448

E-mail [email protected]

www.control3.com

Control Company is ISO 9001 Quality-
Certified by DNV and ISO 17025 accredited
as a Calibration Laboratory by A2LA.

TRACEABLE

®

MEMORY/WATERPROOF THERMOMETER INSTRUCTIONS

Advertising