7. With the desired high temperature alarm 
 
limit on the display, press the SET button (the
current temperature and time-of-day will appear
on the display).
Note:
While in the setting mode, if no button is pressed
for ten (10) seconds, the unit automatically exits from the 
setting mode.
Repeat the above steps to set the second sensor’s 
alarm limits.
VIEWING THE TEMPERATURE ALARM LIMITS
1. Press the S1/S2 button to display the desired
 
temperature sensor.
2. Press the MODE button 3 times, the low
 
temperature alarm limit will appear on the
display (indicated by LO appearing on the
display).
3. Press the MODE button again, the high
 
temperature alarm limit will appear on the
display (indicated by HI appearing on the
display).
4. Press the MODE button a fi nal time to return 
 
to the current temperature/time-of-day display
mode.
Note:
1. If no button is pressed for ten (10) seconds, the unit
 automatically returns to the current temperature/
 time-of-day display mode.
2. The time-of-day displayed when viewing the LO/HI
 alarm limits is the current time-of-day and has no 
 relationship to when an alarm condition was achieved.
 The unit does not record the time-of-day/date that the
 alarm condition was initially achieved.
Repeat the above steps to view the second 
sensor’s alarm limits.
ENABLE/DISABLE ALARMS
To enable the alarm to activate when a temperature 
measured is outside the alarm limits, press and 
release the 
֠
button until “
֠
” appears on the
display. Each successive press of the
֠
button
will enable (
֠
) or disable the alarm.
When the alarm is enabled, both Sensor 1 and 
Sensor 2 alarm set points will be monitored. If either 
sensor reaches an alarm condition, the alarm will 
sound. It is not possible to enable the alarm for one 
sensor and disable the alarm for the other sensor.
ALARM SOUNDING
With the alarm enabled (
֠
), the unit will sound
an alarm and the red LED will fl ash when a 
temperature being measured is outside the alarm 
limits that have been set (equal to or lower than 
the low alarm set point, or equal to or greater than 
the high alarm set point). If the alarm is sounding 
based on the low alarm limit, “LO” will fl ash on the 
display. If the alarm is sounding based on the high 
alarm limit, “HI” will fl ash on the display.
The alarm will sound regardless of which sensor 
is being displayed. Example: If Sensor 1 is being 
displayed, but Sensor 2 reaches an alarm condition, 
the alarm will sound and the display will automatically 
change to display the Sensor 2 temperature.
The unit will continue to alarm and fl ash the 
corresponding “LO” or “HI” until the temperature 
being displayed is within the alarm limits. Once 
the temperature being displayed is within the 
alarm limits, the alarm will stop sounding and the 
corresponding “LO” or “HI” will no longer fl ash; 
however, the red LED will continue to fl ash to 
indicate an alarm condition has occurred. To stop the 
fl ashing LED, the LED must be manually stopped.
To silence an alarm manually, press any button. 
When manually silenced, the corresponding “LO” 
or “HI” will continue to fl ash on the display until the 
temperature is within the alarm limits. The red LED 
will also continue to fl ash until manually stopped.
To manually stop the fl ashing LED, once the tem-
perature has returned to a non-alarm value, press 
and hold the CLEAR button for three (3) seconds.
RESET BUTTON
The RESET button, located on the back of the unit, 
may be used to simultaneously reset the memory 
and alarm limits for both sensors and reset the 
time-of-day and date. Pressing the RESET button 
has the same affect as removing the batteries from 
the unit.
To reset the unit, use a pointed object to press the 
RESET button. When the reset button in pressed, 
the following will occur:
 
• The display go blank.
• The display will show all segments for
approximately 4 seconds.
• The minimum/maximum memory will be
cleared for both sensors.
• The alarm limits for both sensors will be
returned to the factory default value.
• The time-of-day and date will be cleared to
12:00:00 and 01/01/01.
Control Company Cat. No. 4306 / 4307
Traceable
®
is a registered trademark of
Control Company.
©
2009
Control Company.
92-4307-00 Rev. 1 092309
®
BENCH STAND
The unit is supplied with a bench stand that is 
a part of the back of the unit. To use the bench 
stand, locate the small opening at the bottom 
back of the unit. Place your fi ngernail into the 
opening and fl ip the stand out. To close the 
stand, simply snap it shut.
ALL OPERATIONAL DIFFICULTIES
If this unit does not function properly for any 
reason, replace the batteries with new high-quality 
batteries (see the “Battery Replacement” section). 
Low battery power can occasionally cause any 
number of “apparent” operational diffi culties. 
Replacing the batteries with new fresh batteries will 
solve most diffi culties.
BATTERY REPLACEMENT
Erratic readings, a faint display, no display, or a 
battery symbol appearing on the display are all 
indications that the batteries must be replaced. 
Remove the battery cover, located on the back of 
the unit, by sliding it in the direction of the arrow. 
Remove the exhausted batteries and replace them 
with two (2) new AAA alkaline batteries. Make 
certain to insert the new batteries with the proper 
polarity as indicated by the illustration in the battery 
compartment. Replace the battery cover.
Replacing the batteries has the same affect as 
pressing the RESET button. (See the “RESET 
BUTTON” section.)
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281 482-1714 • Fax 281 482-9448
E-mail [email protected] • www.control3.com
Control Company is ISO 9001 Quality-Certifi ed by
DNV and ISO 17025 accredited as a Calibration
Laboratory by A2LA.