Traceable workhorse thermometer instructions
 
SPECIFICATIONS
Range: -58 to 2372 °F
 
-50 to 1300 °C
Resolution: 0.1° or 1° (selectable)
THERMOCOUPLE
The sensor supplied is a Stainless Steel 
Mini-Probe Type-K thermocouple. Maximum 
temperature for this sensor is 932 °F (500 °C), 
with short term usage to 1292 °F (700 °C). 
Accessory thermocouples are available for 
measuring temperatures throughout the entire 
range of the unit (See “Accessories”). Only 
Type-K thermocouples may be used with 
this unit.
Longer leads and extensions may be used. 
The lead effect for 164 feet (50 meters) is 
generally less than 0.2 °C with a limit of error 
of 2.2 °C. Type-K wire and connectors must 
be used.
OPERATION
Insert the thermocouple plug into the socket 
located on the front of the instrument. The 
plug and socket are keyed with a small and 
large pin. Make certain the plug is inserted 
correctly, do not insert the plug incorrectly and 
force it.
Press the POWER button to turn the unit on.
Pressing the °F button will display the tem-
perature in Fahrenheit (°F will appear on the 
display). Pressing the °C button will display 
the temperature in Celsius (°C will appear on 
the display).
Press the 0.1° or 1.0° button to select the 
desired temperature resolution. (Note: 
0.1° resolution can only be used when the 
temperature being measured is between -50° 
and 199.9°)
Place the thermocouple in contact with 
the material to be measured and read the 
temperature on the display.
Press the POWER button to turn the unit off. 
To conserve battery life, always turn the unit 
off when not in use.
HOLD
Press the HOLD button to “freeze” the display 
at the current reading (“H” will appear at the 
top left corner of the display). Press the 
HOLD button again to return to the current 
reading.
OFFSET ADJUSTMENT (Fine tuning)
By adjusting the offset control, found on the 
front of the unit, the measurement accuracy 
may be optimized for a particular thermo-
couple at a particular temperature.
While offset adjustments are possible 
they will VOID with the factory calibration. 
This unit has been calibrated Traceable
®
to
standards provided by the National Institute
of Standards and Technology (NIST). 
Traceable
®
recalibration must be made by
the factory (See the “Warranty, Service or 
Recalibration” section).
BENCH STAND/WALL MOUNT
The unit is supplied with a bench stand that 
is on the back of the case. Locate the stand 
on the back of the unit. With your thumb 
and forefi nger, lift the stand out. To close the 
stand, simply snap it shut.
The unit may be wall mounted by using a 
screw. Set the screw into the wall at the 
desired location. Do not set the screw fl ush 
to the wall, the head of the screw will need 
to slip into the receptacle on the back of the 
unit. Hang the unit by sliding the receptacle 
located on the back of the unit over the head 
of the screw.
©
2006
Control Company. 92-4425-00 Rev. 0
Traceable
®
and Workhorse™ Thermometer are trademarks/
registered trademarks of
Control Company.
Control Company Cat. No. 4425
TRACEABLE
WORKHORSE
THERMOMETER
INSTRUCTIONS
®
™
ACCESSORIES
Control Cat. No. 4008—
Surface Temperature Thermocouple
Control Cat. No. 4014—
Stainless Steel Probe Thermocouple
Control Cat. No. 4028—
Fast Response Beaded Thermocouple
Control Cat. No. 8039—
Low Temperature Probe Thermocouple 
(-400 °F / -240 °C) 
Control Cat. No. 8613—
High Temperature Cable Thermocouple 
(1799 °F / 982 °C)
ALL OPERATIONAL DIFFICULTIES
If this unit does not function properly for any 
reason, replace the battery with a new high 
quality battery (see the “Battery Replacement” 
section). Low battery power can occasionally 
cause any number of “apparent” operational 
diffi culties. Replacing the battery with a new 
fresh battery will solve most diffi culties.
BATTERY REPLACEMENT
Erratic readings, faint readings, no display 
or a battery symbol appearing on the display 
are all indications that the battery should be 
replaced. Before replacing the battery, turn 
the unit off and unplug the probe. Remove the 
battery cover screw located on the back of the 
meter and lift off the battery cover. Remove 
the exhausted battery and replace it with a 
new 9 volt alkaline battery. Replace the 
battery cover and tighten the screw securely. 
WARRANTY, SERVICE, OR RECALIBRATION