Teledyne Analytical Instruments
3 Installation
Model 2020
3-10
Table 3-4: Commands via RS-232 Input
Command
Description
as<enter>
Immediately starts an autospan.
az<enter>
Immediately starts an autozero.
rp<enter>
Allows reprogramming of the
APPLICATION (gas use) and
ALGORITHM (linearization)
System
functions.
st<enter>
Toggling input. Stops/Starts any status message output
from the RS-232, until st<enter> is sent again.
rm1<enter>
Range manual 1
rm2<enter>
Range manual 2
rm3<enter>
Range manual 3
rm4<enter>
Range manual CAL
ra<enter>
Range auto
Implementation: The RS-232 protocol allows some flexibility in its
implementation. Table 3-4 lists certain RS-232 values that are required by
the Model 2020 implementation.
Table 3-5: Required RS-232 Options
Parameter
Setting
Baud
2400
Byte
8 bits
Parity
none
Stop Bits
1
Message Interval
2 seconds
3.3.10 Remote Probe Connector
The 2020 is a single-chassis instrument, which has no Remote Probe
Unit. Instead, the Remote Probe connector is used as another method for
controlling external sample/zero/span gas valves. See Figure 3-6.