4 Operation

Model 2020

4-22

Teledyne Analytical Instruments

4.6.3 Precautions

The Model 2020 allows a great deal of flexibility in choosing ranges

for automatic range switching. However, there are some pitfalls that are to
be avoided.

Ranges that work well together are:

•

Ranges that have the same lower limits but upper limits that
differ by approximately an order of magnitude

•

Ranges whose upper limits coincide with the lower limits of the
next higher range

•

Ranges where there is a gap between the upper limit of the
range and the lower limit of the next higher range.

Range schemes that are to be avoided include:

•

Ranges that overlap

•

Ranges whose limits are entirely within the span of an adjoining
range.

Figure 4-2 illustrates these schemes graphically.

Figure 4-2: Examples of Autoranging Schemes