Model 2230r, process hydrogen analyzer – Teledyne 2230R - Process Hydrogen Analyzer User Manual

Page 13

Advertising
background image

MODEL 2230R, Process Hydrogen Analyzer

Page 13

Relays (Optional) - Connector 2

Supplied Cable – 4m (12 ft.) standard length (Other lengths available).

Wire Color

Description

Grey

Relay 1 Common

Pink

Relay 1 Normally Closed (NC)

Black

Relay 1 Normally Open (NO)

White

Relay 2 Common

Blue

Relay 2 Normally Closed (NC)

Brown

Relay 2 Normally Open (NO)

Serial Interface (Optional) / Remote User Interface Module (Optional) -

Connector 3

Supplied Cable – 4m (12 ft.) standard length (Other lengths available).

Wire

Color

RS232 (standard)

RS422 (Optional)

Description

DB9

Pin

DB25

Pin

Description

DB9

Pin

DB25

Pin

Brown

+6V (N.C.)

White

TxD (Device Transmit)

3

2

TxD- (Device Transmit, -Ve) 3

2

Blue

-

-

-

TxD+ (Device Transmit, +Ve) 4

3

Black

RxD (Device Receive)

2

3

RxD- (Device Receive, -Ve)

2

20

Grey

-

-

-

RxD+ (Device Receive, +Ve) 6

8

Pink

Ground

5

7

Ground

5

7

4.2 ANALOG OUTPUT

The user can request for a specific output current or voltage range or there are a number of

standard analog output ranges the user can select from, which are listed below. The analog

output the user selects is scaled to the user’s hydrogen range of interest. All of this is initially

set at the factory per customer specification at the time of order.

Below is the table for standard analog output current ranges:

Current Analog Output Range

Power-On Self Diagnostic

Error

4 mA to 20 mA

2 mA

3 mA

0 mA to 20 mA

0 mA

20 mA

Advertising