MODEL 2230R, Process Hydrogen Analyzer

Page 19

Now use the 4 BIT-TO-HEXADECIMAL table above to convert this 4 bit value combination into a

two place hexadecimal value:

<fmt> Two Place Hexadecimal Value :

8b

From the OPTIONS <opt> table above, you identify your desired columns with its corresponding

bit value:

<opt> Descriptions

<opt> Bit Value

Calculations Errors

0000 0100

Heater State

0000 0010

<opt> 4 Bit Value Combination: 0000 0110

Again, use the 4 BIT-TO-HEXADECIMAL table to convert this 4 bit value combination into a two

place hexadecimal value:

<opt> Two Place Hexadecimal Value :

06

Conclusion: To have the Time Stamp, Capacitor Reading, Overall Hydrogen Reading, the Peak

Hydrogen Reading, Calculations Errors and Heater State columns continuously reported, you

will implement the “G” serial command (“Go” Command, refer to LEVEL 1 COMMANDS table) as

follows at the H2scan:

command prompt: G < fmt> <opt> = g 8b 06

Display

User Response

H2scan:

Type “g 8b 06”