Teledyne 316RA / RB / RAD / RBD - Oxygen analyzers User Manual
Page 19

TELEDYNE ANALYTICAL INSTRUMENTS
15
Trace Oxygen Analyzer Model 316RA/RB/RAD/RBD
Appendix
Specifications
Ranges:
Standard: 0-10 ppm, 0-100 ppm, 0-1,000
ppm, 0-10,000 ppm, CAL oxygen
Digital: 0-10 ppm, 0-100 ppm, 0-1,000
ppm, 1%, 25%
Sensitivity:
0.5% of full scale
Accuracy:
±2% of full scale at constant temperature
(except ±1 ppm for 0-10 ppm range)
±5% of full scale over the operating
temperature range
(except ±1 ppm for 0-10 ppm range)
Response Time:
At 77°F (25°C):
0-10 ppm, 90% in less than 45 seconds
0-100 ppm, 90% in less than 30 seconds
0-1000 ppm, 90% in less than 10 seconds
0-10,000 ppm, 90 in less than 10 seconds
Operating Temperature:
+32°F to +125°F (0°C to +52°C)
Micro-Fuel Cell Class:
B-2
Micro-Fuel Cell Warranty:
6 months
Voltage Signal Output:
0-1VDC or less (standard)
Current Signal Output:
1-5, 4-20 or 10-50 mADC (optional)
Alarm Output:
(Optional) 1 or 2 DPDT Form C relays
(3A resistive)
Alarm Deadband (Hysteresis): 0.1% of full scale
Power Requirement:
115 VAC, 50/60 Hz, 30 W (100/220
VAC optional)