A-1

Teledyne Analytical Instruments

Portable Oxygen Analyzer

Appendix

Specifications

Standard Ranges

320A, 320B, B/RC: 0-5%, 0-25% and 0-100% Oxygen

BRCD: 0-100% Oxygen

320P: 0-5%, 0-10% and 0-25% Oxygen

Sensitivity

320A, 320B, B/RC, BRCD: 0.5 % of full scale

320P: 1.0% of full scale

Accuracy

320A, 320B, B/RC, BRCD: ±2% of full scale at constant temperature.

±5 % of reading throughout the operating
temperature range, once temperature
equilibrium has been achieved.

320P:

±2 % of full scale or +0.2% oxygen, which-
ever is greater at constant temperature and
pressure of calibration.
±5 % of reading throughout the operating
temperature range (once temperature
equilibrium has been achieved).

Reproducibility

±1% of full scale at constant temperature

Response Time

320A, 320B, B/RC, BRCD: 90% < 10 Seconds (Class B-1)

90% < 15 Seconds (Class B-3)
90% < 32 Seconds (Class C-3)
90% < 45 Seconds (Class A-5)

320P: 90% < 10 Seconds (Class B-1)

NOTE:

Response Time is calculated from sample inlet

of analyzer.

System Operating Temperature

32-122°F (0-50°C)

Appendix