Teledyne 320 Series - Portable oxygen detectors User Manual
Page 15

A-1
Teledyne Analytical Instruments
Portable Oxygen Analyzer
Appendix
Specifications
Standard Ranges
320A, 320B, B/RC: 0-5%, 0-25% and 0-100% Oxygen
BRCD: 0-100% Oxygen
320P: 0-5%, 0-10% and 0-25% Oxygen
Sensitivity
320A, 320B, B/RC, BRCD: 0.5 % of full scale
320P: 1.0% of full scale
Accuracy
320A, 320B, B/RC, BRCD: ±2% of full scale at constant temperature.
±5 % of reading throughout the operating
temperature range, once temperature
equilibrium has been achieved.
320P:
±2 % of full scale or +0.2% oxygen, which-
ever is greater at constant temperature and
pressure of calibration.
±5 % of reading throughout the operating
temperature range (once temperature
equilibrium has been achieved).
Reproducibility
±1% of full scale at constant temperature
Response Time
320A, 320B, B/RC, BRCD: 90% < 10 Seconds (Class B-1)
90% < 15 Seconds (Class B-3)
90% < 32 Seconds (Class C-3)
90% < 45 Seconds (Class A-5)
320P: 90% < 10 Seconds (Class B-1)
NOTE:
Response Time is calculated from sample inlet
of analyzer.
System Operating Temperature
32-122°F (0-50°C)
Appendix