Teledyne 320 Series - Portable oxygen detectors User Manual

Page 8

Advertising
background image

Teledyne Analytical Instruments

2-2

Model 320 Series

Chapter 2

2.3

2.3

2.3

2.3

2.3

Calibration

Calibration

Calibration

Calibration

Calibration

1. Remove the flow-through adapter, and place the range switch in the

0-25% range. On the standard 320P, this is the HI position of the
Range switch.

2. Rotate the Span control until the meter indication pointer coincides

with the red mark extending into the scale of the meter.

3. Repeat this procedure after the cell has been in the instrument at

least 24 hours.

4. After initial calibration, check the span setting before each use.

NOTE: The calibration mark on the 0-25 % range of the detector meter

coincides with the normal oxygen concentration of the atmo-
sphere. Therefore, calibration using this mark as a standard
should not be undertaken in any area where the oxygen content
of the atmosphere surrounding the probe assembly is being
controlled.

2.4

2.4

2.4

2.4

2.4

Micro-Fuel Cell

Micro-Fuel Cell

Micro-Fuel Cell

Micro-Fuel Cell

Micro-Fuel Cell

2.4.1

2.4.1

2.4.1

2.4.1

2.4.1

Cell Replacement

Cell Replacement

Cell Replacement

Cell Replacement

Cell Replacement

When the detector cell nears the end of its useful life, meter readings will
become erratic. Shortly thereafter, the cell output will drop off sharply to
zero. Replace the cell at this point.

NOTE: A spare cell should be ordered soon after receipt of the detector,

and each time the cell is replaced thereafter.

2.4.2

2.4.2

2.4.2

2.4.2

2.4.2

Cell Life and Warranty

Cell Life and Warranty

Cell Life and Warranty

Cell Life and Warranty

Cell Life and Warranty

The MFC carries a warranty that covers its normal life expectancy. How-
ever, far greater life expectancy can be achieved with the cell-saver feature
already discussed. See the specifications section below for warranty,
response time and electrical output of the various cells.

Customers having warranty claims must return the cell in question immedi-
ately to the factory for evaluation. If it is determined that failure is due to
faulty workmanship or parts, the cell will be replaced at no cost to the
customer. Any evidence of tampering, or damage through mishandling, will
render the cell warranty null and void.

If a cell was working satisfactorily, but ceases to function before the war-
ranty period expires, the customer will receive credit, on a prorated basis,
towards the purchase of a new cell.