Part I: 1-5

Percent Paramagnetic Oxygen Analyzer

Part I: Control Unit

Teledyne Analytical Instruments

•

Power Connection

AC power source, 100-240VAC, 50/60
Hz

•

Analog Outputs

0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.

•

Alarm Connections

2 concentration alarms and 1 system
alarm.

•

RS-232 Port

Serial digital concentration signal output
and control input.

•

Remote Probe

Provides all electrical interconnect to the
Analysis Unit or Remote Probe.

•

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration.

•

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

•

Range ID Contacts

Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.

•

Remote Probe

Interfaces with an Analysis Unit or
Remote Probe (external sensor/sample
system).

•

Network I/O

Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3010MB is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.