Teledyne 3010MB - Split architecture paramagnetic oxygen analyzer User Manual
Page 70

A-6
Appendix
Model 3010M
Teledyne Analytical Instruments
Manual Mode Zeroing
ENTER the
ZERO
function. The screen that appears allows you to
select between automatic or manual zero calibration. Use DOWN/UP to
toggle between
AUTO
and
MAN
zero settling. Stop when
MAN
appears on
the display.
Zero: Settling: Man
<ENT> To Begin
ENTER to begin the zero calibration. After a few seconds the first of
five zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predeter-
mined rate. It calculates the differences between successive samplings and
displays the rate of change as
Slope=
a value in parts per million per second
(ppm/s).
#### % Zero
Slope=#### ppm/s
NOTE:It takes several seconds for the true Slope value to display.
Wait about 10 seconds. Then, wait until Slope is sufficiently
close to zero before using
ENTER
to finish zeroing.
Generally, you have a good zero when
Slope
is less than 0.05 ppm/s
for about 30 seconds. When
Slope
is close enough to zero, ENTER it. In a
few seconds, the screen will update.
Once zero settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the ANALYZE
screen.