Teledyne 3010TA - Split architecture trace oxygen analyzer User Manual
Page 96

A-4
Appendix
Model 3010
TA
Diagnostics: Start-up or on-demand, comprehensive, self
testing function initiated by keyboard or
remote command.
Displays: 2 line by 20 alphanumeric, VFD screen, and
one 5 digit LED display. Flowmeter on
Analysis Unit.
Digital Interface: Full duplex RS-232 communications port.
Power: General Purpose Control Unit
Universal power supply 85-250 V ac,
47-63 Hz.
Explosion Proof Analysis Unit
110/220 V ac, 50/60 Hz.
Operating Temperature: 0-50 °C
EMF/RFI: Immunity and Emissions designed to meet
(but not yet certified to)
EN 50081-1
EN 50082-2.
Accuracy: ±2% of full scale at constant temperature.
±5% of full scale over operating temperature
range, on factory default analysis ranges, once
thermal equilibrium has been achieved.
Analog outputs: 0-1 V dc percent-of-range
0-1 V dc range ID.
4-20 mA dc percent-of-range
4-20 mA dc range ID.
Password Access: Can be user-configured for password access.