Teledyne 4040 - Methane / Nonmethane analyzer User Manual

Page 57

Advertising
background image

Operation

Teledyne

Analytical

Instruments

57

Reenter a new Password

'A' 'A' 'A'

________________________________________________

Note that the new password has been replaced by 'A' as placeholder

characters. This is to ensure that the wrong password is not mistakenly
entered.
<▲▼> Use

the

▲▼ keys to reenter the letters of the new

password.

<Enter>

Press <Enter> to advance to the next letter

<Enter>

If the repeated password matches the previously entered
new password, the new password is set, the 4060 resumes
the ANALYSIS SCREEN after displaying a brief
message 'Password Changed'.

<Escape>

Steps back to the previously entered letter or, if on the
first letter, returns to the CHANGE PASSWORD menu.

Note: If you log off the system using the LOGOUT function in

the MAIN MENU, you will now be required to reenter
the password to gain access to most menus.

4.9 Logging Out

The LOGOUT function provides a convenient means of leaving the

analyzer in a password protected mode without having to shut the
instrument off. By entering LOGOUT, you effectively log off the
instrument leaving the system parameters protected until the password is
reentered. The system will continue analysis, however.

To log out:

<▲▼>

From the MAIN MENU scroll to field of LOGOUT
function.

<Enter>

Press <Enter> to logout .
The screen will display the message:

________________________________________________

Restrict Access?

ENTER=Yes ESCAPE=NO

________________________________________________