Teledyne 4060 - Gas Chromatograph User Manual
Page 28
Installation Series
4060
28 Teledyne
Analytical
Instruments
levers found almost at the end of the sliders, both sides at 
the same time. 
Regardless of configuration, the analyzer/system must be installed
on a level surface with sufficient space allocated on either side for 
personnel and test equipment access. Subject to the foregoing, the 
Analyzer/system should be placed as close to the sample point as is 
possible. 
All pertinent dimensions, connecting points, and piping details can
be found in the drawings section as part of the outline, input-output, and 
piping diagrams. These drawings are specific to the instrument or 
system to which the manual applies. 
3.3 User Connections
All user connections are made on the rear panel. Consult the input-
output and outline diagrams in the drawing section of the manual. Not 
all the features displayed may be present in your system. Refer to any 
Addenda for additional information that may apply to your instrument. 
3.3.1 Electrical Power Connections
The standard analyzer requires a supply of 100-125VAC, single-
phase power. Power connections are made at the rear panel of the unit. 
Refer to the input-output diagram for more information. The electrical 
power service must include a high-quality ground wire. A high-quality 
ground wire is a wire that has zero potential difference when measured 
to the power line neutral. If you have the 220 VAC option, you will 
require 220 or 240 VAC, 50/60 Hz power. Check the analyzer input-
output diagram, power schematic, outline, and wiring diagrams for 
incoming power specifications and connecting points. 
CAUTION:
PRIMARY POWER TO THE SYSTEM SHOULD NOT 
BE SUPPLIED UNTIL ALL CUSTOMERS WIRING IS 
INSPECTED PROPERLY BY START-UP PERSONNEL. 
3.3.2 Electronic Connections
Figure 3-1 shows the Series 4060 rear panel. There are connections
for power, digital communications, and both digital and analog 
concentration output.