3. External hazardous voltages which may be
connected to the Analyzer alarm relay 
connections. 
Procedure for Removal of Internal Inaccessible 
Shock Hazards 
CAUTION:
SERVICING OR MAINTENANCE OF THE ANALYZER 
SHOULD ONLY BE DONE BY SUITABLE TRAINED 
PERSONNEL. TO AVOID THESE INACCESSIBLE 
HAZARDOUS VOLTAGES WHEN SERVICING THE 
SERIES 4060, PERFORM EACH OF THE FOLLOWING 
STEPS, IN THE ORDER GIVEN, BEFORE SERVICING 
BEGINS: 
1. Switch off the power to the Analyzer and remove the main (line)
power cord from the Analyzer.
2. Remove all external voltages from the connections to the alarm
contacts.
3. Wait one minute.
4. Discharge the anode supply voltage.
a. Connect one end of an insulated (to 1000 VDC or more)
clip lead to Analyzer chassis ground (the standoff for the 
upper right corner of the mother PCB). 
b. Put one end of a 500V rated 1000 ohm resistor in the
other end of the clip lead.
c. Check the voltage between chassis ground (the standoff
for the upper right corner of the mother PCB) and the top 
side of R2 at PCB number B74671. It should be between 
-5VDC and +5VDC. If is in that range, the inaccessible 
hazardous voltage removal procedure is completed, if not 
repeat steps 4.a and 4.b. 
If it is absolutely necessary to work inside the instrument with power
on, use the ONE HAND RULE:
Work with one hand only.
Keep the other hand free without contacting any other object. This
reduces the possibility of a ground path through the body in case of 
accidental contact with hazardous voltages.