Teledyne 6000A - Photometric analyzer User Manual

Page 48

Advertising
background image

3 Operation

Model 6000A

3-24 Part I

Teledyne Analytical Instruments

If above screen displays, use the

∆∇ arrow keys to Select AUTO, and

press

Enter.

Press

Escape

to return to the previous Analyze Function.

3.6.3 Precautions

The Model 6000A allows a great deal of flexibility in choosing ranges for

automatic range switching. However, there are some pitfalls that are to be
avoided.

Ranges that work well together are:

•

Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude

•

Ranges whose upper limits coincide with the lower limits of the
next higher range

•

Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.

Range schemes that are to be avoided include:

•

Ranges that overlap

•

Ranges whose limits are entirely within the span of an adjoining
range.

•

Ranges where the zero is suppressed, is 1-10, 1-100, etc,
however, 80-100, 90-100 is ok where the zero gas is actually
100% concentration and the calibration is inverted.

Figure 3-2 illustrates these schemes graphically.