Teledyne 7120 - NDIR analyzer User Manual

Page 56

Advertising
background image

Infrared Analyzer

Operation 4

4-23

Teledyne Analytical Instruments

the number. Holding down the key speeds up the incrementing or
decrementing.

•

After the number (value) has been choosed, use Enter to move
the desired parameter. Then use the UP/DOWN keys to change
the parameter.

•

Once the parameters for alarm have been set, Enter the alarm
function again, and repeat this procedure for next alarm.

•

To reset a latched alarm, go to

Dft– and then use either UP two

times or DOWN two times. (Toggle it to

Y and then back to N.)

–OR –

Go to

Ltch– and then use either UP two times or DOWN two

times. (Toggle it to

N and back to Y.)

4.6

The Range Select Function

The Range function allows you to manually select the concentration range

of analysis (

MANUAL), or to select automatic range switching (AUTO).

In the

MANUAL screen, you are further allowed to define the high and low

(concentration) limits of each Range, and select a single, fixed range to run.

CAUTION: If this is a linearized application, the new range must

be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.

In the

AUTO screen, you are further allowed to select which gas application

(PREVIOUSLY defined in APPLICATION function) to run.

4.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for the
range.

Use UP/DOWN key to start the RANGE function, and ENTER

Select range
mode: MANUAL