ZIRCONIA OXYGEN SENSOR OUTPUT (mV)
PROBE TYPE 1231, SENSOR TYPE 1234
% OXYGEN
mV at 720°C (1320°F)
___________________________________________________
21.0
0.00
20.5
0.46
20.0
0.99
19.5
1.53
19.0
2.09
18.5
2.66
18.0
3.25
17.5
3.85
17.0
4.47
16.5
5.11
16.0
5.77
15.5
6.45
15.0
7.15
14.5
7.87
14.0
8.62
13.5
9.40
13.0
10.21
12.5
11.05
12.0
11.92
11.5
12.83
11.0
13.78
10.5
14.78
10.0
15.82
9.5
16.92
9.0
18.08
8.5
19.30
8.0
20.60
7.5
21.98
7.0
23.45
6.5
25.04
6.0
26.75
5.5
28.61
5.0
30.65
4.5
32.90
4.0
35.42
3.5
38.28
3.0
41.58
2.5
45.48
2.0
50.25
1.5
56.41
1.0
65.08
0.5
79.91
0.2
99.51
_________________________________
'K’ TC mV
29.212 at 720
°C (1320°F)
These tables are based on the Nernst equation:
Sensor e.m.f. = 0.02154 x T x 1n x 20.95 / % oxygen, where T = ° K (° C + 273), e.m.f. is in mV’s