Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual
Page 90

9060-Zirconia
Analyser
Configuration and Setup Addresses
Holding
Reg. Function
Description
716 Probe #1 offset
10 = 1.0mV
717 Probe #2 offset
10 = 1.0mV
Purge control related variables
754 Purge enable
0= off, 1= on
Calibration checking gas related variables
759 Gas calibration check
0= off, 1= 1 gas, 2= 2 gasses
204
8
Probe #1 EMF
100,000 = 100.000 mV
205
0
Probe #2 EMF
100,000 = 100.000 mV
205
2
Probe #1 OXYGEN
100,000,000 = 100.0%
205
4
Probe #2 OXYGEN
100,000,000 = 100.0%
205
6
Probe #1, Impedance
1,000 = 1 k
Ω
205
8
Probe #2, Impedance
1,000 = 1 k
Ω
206
0
Probe #1 TC mV
100,000 = 100.000 mV
206
2
Probe #2 TC mV
100,000 = 100.000 mV
206
4
Probe #1 temperature
700 = 700 degC
206
6
Probe #2 temperature
700 = 700 degC
206
8
ALRM-ARRAY
Array of current alarm status. See below
208
4
ALRM-TIMES
Array of timestamp of alarms
Alarm array order -
1. Heater 1 fail
2. Sensor1 fail (Impedance too high)
3. Probe 1 filter blocked
4. Probe 1 thermocouple open circuit
5. Reference air Pump fail
6. Battery backed RAM fail
7. Mains frequency measurement fail
8. ADC warning (outside normal specifications, but still accurate)
9. DAC warning (outside normal specifications, but still accurate)