Operational Theory

BDS 3960

Teledyne Analytical Instruments

22

2.2.3 Gas

Pressure

The analyzer is equipped with a pressure regulator as shown in

Figure 2-3. To access the pressure regulator, the front panel must be
opened. The inlet pressure should be regulated anywhere between 4 to
50 psig. The sensor is not affected by pressure changes in the inlet as
long as the analyzer vents to atmosphere. If the analyzer is not vented to
atmosphere, the downstream pressure must not exceed 10 inch of water.
A clogged or restricted vent or excessive pressure will force gas into the
electrolyte and cause damage to the BDS sensor.

Pressure Regulator

BDS Sensor

Flowmeter

Vent

Oxygen Scrubber

Sample

Inlet

Span

Inlet

Figure 2-3: A simplified BDS Sample System

2.2.4 Temperature

effect

The raw output from a BDS oxygen sensor has a temperature

coefficient about 0.25% / °C. That is an average value, it changes as
temperature changes. This temperature effect is compensated by the
software throughout the operation temperature range (5 – 40°C).

2.2.5

Recovery from High Level Oxygen Exposure

The ambient air contains about 210,000,000 ppb (2.1 x 10

8

)

oxygen. Figure 2.4 is a typical purge-down curve for a new BDS sensor
which had been air saturated. It is normal to take several hours, even
days for an air saturated BDS to purge down to a low ppb level.