Teledyne BDS-3960 - Oxygen analyzer User Manual

Page 35

Advertising
background image

Oxygen Analyzer

Installation

Teledyne Analytical Instruments

35

4–20 mA DC % Range:

Current increases linearly with
increasing oxygen, from 4 mA at 0
ppm to 20 mA at full scale ppm. (Full
scale = 100% of programmable range.)

4–20 mA dc Range ID:

8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range, 20 mA
= 100ppm.

Table 3-1: Analog Output Connections

Pin

Function

3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ % Range, 4-20 mA, floating

6

– % Range, 4-20 mA, floating

8

+ Range ID, 0-1 V dc

23

– Range ID, 0-1 V dc, negative ground

24

+ % Range, 0-1 V dc

7

– % Range, 0-1 V dc, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The connectors are:

Threshold Alarm 1:

•

Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).

•

Can be configured as failsafe or nonfailsafe.

•

Can be configured as latching or nonlatching.

•

Can be configured out (defeated).