Teledyne BDS-3960 - Oxygen analyzer User Manual
Page 35

Oxygen Analyzer
Installation
Teledyne Analytical Instruments
35
4–20 mA DC % Range:
Current increases linearly with
increasing oxygen, from 4 mA at 0
ppm to 20 mA at full scale ppm. (Full
scale = 100% of programmable range.)
4–20 mA dc Range ID:
8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range, 20 mA
= 100ppm.
Table 3-1: Analog Output Connections
Pin
Function
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ % Range, 4-20 mA, floating
6
– % Range, 4-20 mA, floating
8
+ Range ID, 0-1 V dc
23
– Range ID, 0-1 V dc, negative ground
24
+ % Range, 0-1 V dc
7
– % Range, 0-1 V dc, negative ground
Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The connectors are:
Threshold Alarm 1:
•
Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).
•
Can be configured as failsafe or nonfailsafe.
•
Can be configured as latching or nonlatching.
•
Can be configured out (defeated).